FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2010-00994
- Event Type
- Malfunction
- Date Received
- July 27, 2010
- Date of Event
- July 3, 2010
- Report Date
- August 24, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
THIS IS A FINAL REPORT. THE CUSTOMER RETURNED METER SERIAL NUMBER (B)(4). THE METER HAS BEEN TESTED WITH AN APPROVED STRIP LOT 1002834. THE COMPLAINT WAS NOT CONFIRMED AND ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. IN ADDITION, THE READINGS REPORTED WERE NOT FOUND IN THE INTERNAL MEMORY LOG OF THE METER.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 429 MG/DL AND 129 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
REPORTER ALLEGED THAT HE OBTAINED A RESULT OF 113 MG/DL ON THE ADVANTAGE SYSTEM WITH EXPIRED STRIPS AND FELT CONFUSED BEFORE PASSING OUT 15 MINUTES LATER. THE CUSTOMER REPORTED THAT HIS FRIENDS CALLED AN EMERGENCY UNIT, THEY ARRIVED 15 MINUTES LATER, OBTAINED A RESULT OF 25 MG/DL ON THEIR SYSTEM AND TREATED THE CUSTOMER WITH MEDICATION. REPORTER STATED THAT HE WAS TAKEN TO THE HOSPITAL AND GIVEN ORANGE JUICE AND FOOD THERE. NO OTHER ACTIONS WERE REPORTED TAKEN OR TREATMENT RECEIVED. A REQUEST WAS MADE FOR THE RETURN OF THE SUSPECT DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1011502 |