ID NOW COVID-19 2.0 TEST KIT 24T JAPAN
Report
- Report Number
- 1221359-2023-01397
- Event Type
- Malfunction
- Date Received
- September 7, 2023
- Date of Event
- August 26, 2023
- Report Date
- October 20, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D4-UDI:((B)(6) THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 192-000J THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M232058 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 192-000J/ LOT: M232058, TEST BASE PART NUMBER 192-430 / LOT: M232058. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M232058 SHOWED THAT THE COMPLAINT RATE IS(B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED
D4-UDI:(B)(4). THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 192-000J THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 192-000. THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON(B)(6)2023. CONFIRMATION TESTING WAS PERFORMED ON THE SAME DAY VIA PCR (PLATFORM: UNKNOWN) AND GENERATED A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 2.0 ASSAY PERFORMED ON (B)(6) 2023. CONFIRMATION TESTING WAS PERFORMED ON THE SAME DAY VIA PCR (PLATFORM: UNKNOWN) AND GENERATED A NEGATIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1948281 | ID NOW COVID-19 2.0 TEST KIT 24T JAPAN | REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC ACID | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | M232058 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |