BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE
Report
- Report Number
- 1911916-2023-00643
- Event Type
- Malfunction
- Date Received
- September 7, 2023
- Date of Event
- August 21, 2023
- Report Date
- December 11, 2023
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FOZ
- UDI-DI
- 30382903065463
- PMA / PMN Number
- K003553
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 306546 AND LOT NUMBER 3103219. THE REVIEW DID NOT REVEAL ANY POSSIBLE NON-CONFORMANCES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS ISSUE, 132 PHYSICAL SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THE SAMPLES WERE ALL LOOSELY PACKED IN A BOX WITH THEIR ORIGINAL FLOW WRAP PACKAGING. GLIDE FORCE TESTING WAS PERFORMED ON ALL OF THE RETURNED SAMPLES. OUT OF 132 SAMPLES TESTED, 17 FAILED THE GLIDE FORCE TESTING; CONFIRMING THE REPORTED DEFECT OF PLUNGER MOVEMENT DIFFICULTY DUE TO DRY BARRELS. THE MOST PROBABLE CAUSE FOR THE DEFECTIVE PRODUCT WAS A FAULTY THERMOCOUPLE, WHICH WAS REPLACED EARLIER THIS YEAR. THE THERMOCOUPLE CONTROLS THE DISPERSION OF SILICONE IN THE SYRINGE BARREL. THE FAULTY THERMOCOUPLE IN TURN LED TO INTERMITTENT DEFECTS WHICH WERE NOT DETECTED DURING THE IN-PROCESS CHECKS. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.
H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE PLUNGER ON THE BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE WAS DIFFICULT TO MOVE. THE FOLLOWING WAS RECEIVED BY THE INITIAL REPORTER: CUSTOMER HAVING TROUBLE PUSHING OUT FLUIDS 10 ML PREFILLED SYRINGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1038349 | BD POSIFLUSH¿ PREFILLED NORMAL SALINE FLUSH SYRINGE | SALINE, VASCULAR ACCESS FLUSH | FOZ | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 3103219 | 30382903065463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |