FDA Adverse Event Malfunction Summary report: N

CLICKFINE 32GX4MM DONGBAO 7CT

MDR report key: 17701945 · Received September 7, 2023

Report

Report Number
3002806818-2023-00010
Event Type
Malfunction
Date Received
September 7, 2023
Date of Event
June 20, 2023
Report Date
September 7, 2023
Manufacturer
YPSOMED AG
Product Code
FMI
PMA / PMN Number
K171984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE HOSPITAL'S PHARMACY OBSERVED A DAMAGED PEEL FOIL, A STERILE BARRIER, PRIOR TO USE. THE REVIEW OF MANUFACTURING HISTORY SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FROM THE MAIN BATCH 221225. NO ABNORMALITIES WERE FOUND ON THE PEEL FOIL OR SEAL SEAM OF RETAINED SAMPLES OF THE SAME BATCH. THE PRODUCT WAS NOT USED ON A PATIENT.

Description of Event or Problem · 0

THE HOSPITAL'S PHARMACY OBSERVED A DAMAGED PEEL FOIL, A STERILE BARRIER, PRIOR TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1100713 CLICKFINE 32GX4MM DONGBAO 7CT PEN NEEDLE FMI YPSOMED AG CLICKFINE 32GX4MM DONGBAO 7CT 211225-02

Patients

Seq Age Sex Outcome Treatment
1 Unknown