FDA Adverse Event
Malfunction
Summary report: N
CLICKFINE 32GX4MM DONGBAO 7CT
MDR report key: 17701945
·
Received September 7, 2023
Report
- Report Number
- 3002806818-2023-00010
- Event Type
- Malfunction
- Date Received
- September 7, 2023
- Date of Event
- June 20, 2023
- Report Date
- September 7, 2023
- Manufacturer
- YPSOMED AG
- Product Code
- FMI
- PMA / PMN Number
- K171984
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THE HOSPITAL'S PHARMACY OBSERVED A DAMAGED PEEL FOIL, A STERILE BARRIER, PRIOR TO USE. THE REVIEW OF MANUFACTURING HISTORY SHOWED NO ABNORMALITIES OR DEVIATIONS FROM THE VALIDATED MANUFACTURING PROCESS FROM THE MAIN BATCH 221225. NO ABNORMALITIES WERE FOUND ON THE PEEL FOIL OR SEAL SEAM OF RETAINED SAMPLES OF THE SAME BATCH. THE PRODUCT WAS NOT USED ON A PATIENT.
Description of Event or Problem · 0
THE HOSPITAL'S PHARMACY OBSERVED A DAMAGED PEEL FOIL, A STERILE BARRIER, PRIOR TO USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1100713 | CLICKFINE 32GX4MM DONGBAO 7CT | PEN NEEDLE | FMI | YPSOMED AG | CLICKFINE 32GX4MM DONGBAO 7CT | 211225-02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|