NURSE CALL
Report
- Report Number
- 2027454-2023-00074
- Event Type
- Malfunction
- Date Received
- September 7, 2023
- Date of Event
- August 15, 2023
- Report Date
- September 7, 2023
- Manufacturer
- HILL-ROM CARY
- Product Code
- ILQ
- UDI-DI
- 00887761985209
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE HILLROM TECHNICIAN FOUND THE CODE BLUE WAS NOT CONFIGURED FOR SOME ROOMS. THE NAVICARE® NURSE CALL¿ (NNC) SYSTEM IS A COMPREHENSIVE COMMUNICATION AND INFORMATION MANAGEMENT SYSTEM THAT HELPS CAREGIVERS PROVIDE THE BEST POSSIBLE CARE TO PATIENTS IN THE MOST EFFICIENT MANNER. THE SYSTEM FACILITATES COMMUNICATION WITHIN A NURSING UNIT. CAREGIVERS HAVE THE ABILITY TO QUICKLY LOCATE AND COMMUNICATE WITH A FELLOW STAFF MEMBER, AND PATIENTS CAN EASILY COMMUNICATE WITH CAREGIVERS, THE NURSING STATION, OR DIRECTLY TO THEIR PRIMARY CAREGIVER. THE HRC TECHNICIAN CONFIGURED PBX TO MONITOR CODE BLUE IN UNIT FOR THE CUSTOMER TO RESOLVE THE ISSUE. ALTHOUGH THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT, IF THE REPORTED ISSUE WERE TO RECUR IT COULD LEAD TO INJURY OR DEATH. THEREFORE HILLROM IS REPORTING THIS COMPLAINT.
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE CODE BLUE WAS NOT SHOWING IN THE PBX. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 925732 | NURSE CALL | SYSTEM, COMMUNICATION, POWERED | ILQ | HILL-ROM CARY | 00887761985209 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |