Narratives
THIS ISSUE IS DEEMED A REPORTABLE EVENT SINCE THE MALFUNCTION "FLOW SENSOR CALIBRATION FAILS" CAN LEAD TO A DELAY IN THE THERAPY SINCE THE VENTILATOR CANNOT BE USED. ANOTHER VENTILATOR MUST BE MADE AVAILABLE. THE ROOT CAUSE CANNOT BE DETERMINED. WITHIN THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT. THERE WAS NO PATIENT OR USER HARM. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE.
"TIGHTNESS TEST" AND "FLOW SENSOR CALIBRATION" ARE NG. WHEN I CHECKED ON (B)(6) ONLY "FLOW SENSOR CALIBRATION" WAS NG. WHEN I CHECKED AGAIN TODAY, BOTH WERE NG. I HAVE CHECKED WITH MULTIPLE CIRCUITS AND THERE IS A REPRODUCTION.