AMIS 01.15S.100 AMIS LEG POSITIONER 2.0 FRAME

MDR report key: 17697148 · Received September 7, 2023

Report

Report Number
3005180920-2023-00678
Event Type
Injury
Date Received
September 7, 2023
Date of Event
August 8, 2023
Report Date
September 7, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
FWX
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON (B)(6) 2023. LOT 2155343: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 01-FEB-2022. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THE SAME LOT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY MEDICA AFFAIRS DEPARTMENT: DURING PRIMARY THA, THE GREATER TROCHANTER WAS FRACTURED DURING THE EXTERNAL ROTATION WITH THE LEG POSITIONER. ACCORDING TO REPORT, THE BONE QUALITY OF THE PATIENT WAS WEAK, AND ALSO FROM THE RADIOGRAPHIC IMAGE THIS CAN BE CONFIRMED. IN ORDER TO FIX THE FRACTURE, PLATE AND CABLES WERE PLACED. THERE IS NO REASON TO SUSPECT A FAULTY DEVICE. OTHER DEVICES INVOLVED: STEM: AMISTEM C 01.18.700 AMISTEM-C SHORT NECK STD STEM SIZE 0 LOT. 2240295 LOT 2240295: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 09-DEC-2022. EXPIRATION DATE: 2027-NOV-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. AMISTEM-C SN IS NOT REGISTERED IN USA.

Description of Event or Problem · 0

THE GREATER TROCHANTER WAS FRACTURED DURING THE EXTERNAL ROTATION WITH THE LEG POSITIONER. PLATE AND CABLES WERE PLACED TO FIX THE FRACTURE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE BONE QUALITY WAS WEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
926315 AMIS 01.15S.100 AMIS LEG POSITIONER 2.0 FRAME HIP LEG POSITIONER FWX MEDACTA INTERNATIONAL SA 2155343

Patients

Seq Age Sex Outcome Treatment
1 Female Other