PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2023-09679
- Event Type
- Injury
- Date Received
- September 6, 2023
- Date of Event
- August 21, 2023
- Report Date
- October 18, 2023
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- UDI-DI
- 08717648013089
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED CUFF MISS COULD NOT BE TESTED DUE TO DEVICE COMPONENTS NOT BEING RETURNED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.D4 - LOT # UPDATED FROM 3061041 TO 3061541. H4: MANUFACTURING DATE CORRECTED AS A RESULT.
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
IT WAS REPORTED THAT A VENOUS CLOSURE OF THE RIGHT COMMON FEMORAL VEIN WAS ATTEMPTED WITH A PROGLIDE DEVICE AFTER A SUPRAVENTRICULAR TACHYCARDIA INTERVENTIONAL PROCEDURE USING AN 8.5F SHEATH. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED. A NEW PROSTYLE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1006858 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 3061541 | 08717648013089 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |