FDA Adverse Event Injury Summary report: N

MONOJCT 20ML SYR L-LOCK

MDR report key: 17690373 · Received September 6, 2023

Report

Report Number
1282497-2023-10780
Event Type
Injury
Date Received
September 6, 2023
Date of Event
August 25, 2023
Report Date
February 12, 2024
Manufacturer
CARDINAL HEALTH
Product Code
FMF
UDI-DI
10884521010802
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE HAS BEEN CONFIRMED AND A CAPA-CRP-03163 HAS BEEN INITIATED TO FURTHER INVESTIGATE. IN ADDITION, THE REPORTED DEVICE IS PART OF A RECALL, EVENT-2023-05729/RES 93075. PLEASE REFER TO THE RECALL FOR ADDITIONAL DETAILS.

Additional Manufacturer Narrative · 0

AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.

Additional Manufacturer Narrative · 0

H6 MEDICAL DEVICE PROBLEM CODE WAS UPDATED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THEY ARE USING THE CARDINAL HEALTH (CAH) BRAND SYRINGES IN THE ALARIS INFUSION PUMP AND EXPERIENCING MEDICATION DELIVERING FASTER THAN IT SHOULD. FOR EXAMPLE, IF THE PUMP IS SET TO INFUSE OVER 13 HOURS, IT IS DELIVERING MEDICATION IN 6-7 HOURS AS IT IS DETECTING INCORRECT VOLUMES IN THE PUMPS. EXAMPLE: THE PUMP IS DETECTING A VOLUME OF 7, BUT THERE IS ONLY A VOLUME OF 5, SO IT IS ADMINISTERING FASTER THAN SET AS THE PUMP BELIEVES THERE IS MORE MEDICATION TO DELIVER OVER THE PERIOD OF TIME. THIS HAS LED TO PATIENTS RECEIVING "LIFE SUSTAINING" MEDICATION TOO QUICKLY. THE CUSTOMER STATED THE PHARMACY DID THEIR OWN TESTING USING WATER IN THE CAH BRAND 3ML, 6ML, 20ML AND 35ML SYRINGES IN THE ALARIS PUMP AND EACH TIME THE PUMP DETECTED INCORRECT VOLUME, STATING 100% FAILURE RATE. THE HOSPITAL IS PULLING ALL CAH BRAND SYRINGES FROM THE FLOORS, STARTING WITH CRITICAL AREAS, AND WILL SEQUESTER PRODUCT FOR INVESTIGATION SAMPLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1056901 MONOJCT 20ML SYR L-LOCK SYRINGE, PISTON FMF CARDINAL HEALTH 1182000777 10884521010802

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other