30K FSI-SLI-10S INSERT,PKD
Report
- Report Number
- 2424472-2023-00406
- Event Type
- Malfunction
- Date Received
- September 6, 2023
- Report Date
- December 1, 2023
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- UDI-DI
- D003803951
- PMA / PMN Number
- K052334/K970
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
FAILURE MODE: OVERHEATING INSERT ROOT CAUSE: TIP/NOZZLE CLOGGED (QTY. 2) CONCLUSION CODE: UNEXPECTED FAILURE RETURNED QTY.2 INSERTS, 1. 80395, 22230-00081527, JX, BUL TEST METHOD / (B)(4) INDUCTANCE: GAUGE ID# 5349 DUE: 11/30/2024 RESULT: 3.17 MEETS SPEC DOES NOT MEET SPEC N/A TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC N/A STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A TESTED ON: G136 GAGE ID# 9626-2 DUE DATE: 03/31/24 SET PRESSURE: 35 PSI WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A SPRAY PATTERN: - MEETS SPEC DOES NOT MEET SPEC N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC N/A GRIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A OTHER VISUAL OBSERVATION: INSERT TIP IS CLOGGED, NO WATER FLOW. ALLEGED EVENT: CONFIRMED - NOT CONFIRMED 2. 80395, 22230-00081527, JX, BUL TEST METHOD / (B)(4) INDUCTANCE: GAUGE ID# 5349 DUE: 11/30/2024 RESULT: 3.21 MEETS SPEC DOES NOT MEET SPEC N/A TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC N/A STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A TESTED ON: G136 GAGE ID# 9626-2 DUE DATE: 03/31/24 SET PRESSURE: 35 PSI WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A SPRAY PATTERN: - MEETS SPEC DOES NOT MEET SPEC N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC N/A GRIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A OTHER VISUAL OBSERVATION: INSERT TIP IS CLOGGED, NO WATER FLOW. ALLEGED EVENT: CONFIRMED - NOT CONFIRMED.
SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803.THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT IT IS REPORTED THAT 2 30K FSI-SLI-10S INSERT,PKD ARE NOT VIBRATING, HAVE NO WATER FLOW AND ARE GETTING HOT DURING USE. NO INJURY OCCURRED.