FDA Adverse Event Malfunction Summary report: N

30K FSI-SLI-10S INSERT,PKD

MDR report key: 17689875 · Received September 6, 2023

Report

Report Number
2424472-2023-00406
Event Type
Malfunction
Date Received
September 6, 2023
Report Date
December 1, 2023
Manufacturer
DENTSPLY LLC
Product Code
ELC
UDI-DI
D003803951
PMA / PMN Number
K052334/K970
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

FAILURE MODE: OVERHEATING INSERT ROOT CAUSE: TIP/NOZZLE CLOGGED (QTY. 2) CONCLUSION CODE: UNEXPECTED FAILURE RETURNED QTY.2 INSERTS, 1. 80395, 22230-00081527, JX, BUL TEST METHOD / (B)(4) INDUCTANCE: GAUGE ID# 5349 DUE: 11/30/2024 RESULT: 3.17 MEETS SPEC DOES NOT MEET SPEC N/A TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC N/A STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A TESTED ON: G136 GAGE ID# 9626-2 DUE DATE: 03/31/24 SET PRESSURE: 35 PSI WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A SPRAY PATTERN: - MEETS SPEC DOES NOT MEET SPEC N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC N/A GRIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A OTHER VISUAL OBSERVATION: INSERT TIP IS CLOGGED, NO WATER FLOW. ALLEGED EVENT: CONFIRMED - NOT CONFIRMED 2. 80395, 22230-00081527, JX, BUL TEST METHOD / (B)(4) INDUCTANCE: GAUGE ID# 5349 DUE: 11/30/2024 RESULT: 3.21 MEETS SPEC DOES NOT MEET SPEC N/A TIP CONDITION: - MEETS SPEC DOES NOT MEET SPEC N/A STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A TESTED ON: G136 GAGE ID# 9626-2 DUE DATE: 03/31/24 SET PRESSURE: 35 PSI WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A SPRAY PATTERN: - MEETS SPEC DOES NOT MEET SPEC N/A WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING SEAM LEAK N/A VIBRATION: - MEETS SPEC DOES NOT MEET SPEC N/A GRIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A OTHER VISUAL OBSERVATION: INSERT TIP IS CLOGGED, NO WATER FLOW. ALLEGED EVENT: CONFIRMED - NOT CONFIRMED.

Additional Manufacturer Narrative · 0

SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803.THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT 2 30K FSI-SLI-10S INSERT,PKD ARE NOT VIBRATING, HAVE NO WATER FLOW AND ARE GETTING HOT DURING USE. NO INJURY OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1831610 30K FSI-SLI-10S INSERT,PKD SCALER, ULTRASONIC ELC DENTSPLY LLC 22230/JX D003803951

Patients

Seq Age Sex Outcome Treatment
1 Unknown