HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-20451
- Event Type
- Malfunction
- Date Received
- September 5, 2023
- Date of Event
- August 29, 2021
- Report Date
- September 4, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K193228
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG MORE THAN 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS USED. THE ROOT CAUSE WAS DETERMINED TO BE A COMMUNICATION FAILURE BETWEEN VU ESM BOARD AND IP ESM BOARD. IN CONSEQUENCE (CORRECTION) THE VU ESM BOARD AND IP ESM BOARD WERE REPLACED. THERE WAS NO PATIENT OR USER HARM.
HELLO, BELOW IS THE TEXT FROM EMAILS FROM THE RT AND BIOMED AT UF SHANDS WITH "PANEL DISCONNECT" . WILL BE ONSITE TO REPLACE VU & IP ESM'S ASAP. ---"THIS IS VENT THAT CRASHED ON PT. I ATTACHED EVENT LOGS. RT SAID CONTROL PANEL DISCONNECT ON PATIENT. NO HARM TO PATIENT. VENT WILL NOT TURN OFF. (SIGMA#139043 OUR ASSET NUMBER). THIS HAPPENED AT (B)(6) 2021 AT 3AM. BELOW IS RT JUSTIN GILMORES NOTES. WHEN I BOOTED IT UP THIS MORNING, IT BOOTED UP AS IF IT WAS STILL RUNNING ON PATIENT. I PLACED INTO STANDBY MODE AND FOUND PANEL DISCONNECT, BUT ALSO MANY LOSS OF AIR AND O2 ALARMS AND TIME AND DATE WAS WRONG. EVERYTHING POINTED TO A LOSS OF COMMUNICATION BETWEEN UI AND BDU. BATTERY WAS EMPTY, BUT THAT WAS PROBABLY THEM UNPLUGGING VENT TO STOP ALARMING. VENT RECORDED PANEL DISCONNECT AT (B)(6) 2021 AT 9AM, BUT LOGS THEN SKIP FROM (B)(6) 2021, TO (B)(6) 2021 THEN TO (B)(6)2020 ON A BRAND NEW VENTILATOR. HEY, I JUST WANTED TO GIVE YOU GUYS A HEADS UP ON A HAMILTON G5 THAT WE HAD ATTACHED TO A PATIENT IN UA4222. AT 300 AM THE THERAPIST NOTICED THE VENT ALARMING. ONCE SHE WENT INTO THE ROOM SHE NOTICE IT SAID PANEL CONNECTION LOST . THE VENT WAS NOT VENTILATING THE PATIENT SO SHE UN-HOOKED PATIENT AND STARTED BAGGING. NO HARM WAS DONE TO THE PATIENT. WE DID PLACE A PSR ON THE VENT. I ALSO PLACED A WORK ORDER IN AND PLACED IT IN COLINS OFFICE WITH THE BROKEN EQUIPMENT FORM ON IT. SIGMA #139043. LET ME KNOW IF THERE IS ANYTHING I CAN DO TO HELP. THANKS (B)(6)"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1639434 | HAMILTON MEDICAL AG | HAMILTON-G5 | CBK | HAMILTON MEDICAL AG | HAMILTON-G5 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |