EVIS EXERA III VIDEO SYSTEM CENTER
Report
- Report Number
- 3002808148-2023-09351
- Event Type
- Malfunction
- Date Received
- September 5, 2023
- Date of Event
- August 14, 2023
- Report Date
- November 16, 2023
- Manufacturer
- SHIRAKAWA OLYMPUS CO., LTD.
- Product Code
- FET
- UDI-DI
- 04953170298622
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. ADDITIONAL INFORMATION ADDED TO FIELD H6. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE FACTS OBTAINED FROM THE INVESTIGATION, IT IS PRESUMED THAT [NO IMAGE] OCCURRED DUE TO WRONG INPUT SELECTION. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IN SPEAKING WITH THE OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC), THE CUSTOMER REPORTED THE EVIS EXERA III VIDEO SYSTEM CENTER IS NOT DISPLAYING AN IMAGE ON THE OEV-191H / OEV 262H MONITORS. THROUGH TROUBLESHOOTING, THE CABLING WAS VERIFIED: SERIAL DIGITAL INTERFACE (SDI) OUT 1 GOES TO OEV-191H VIDEO IN PORT (CANNOT SELECT SDI ON THE OEV-191H), COMP OUT GOES TO GMED, THE MONITOR OUT GOES TO THE WALL, THE OEV-262H HAS A CABLE CONNECTED TO THE VIDEO IN, AND THERE IS A RED, GREEN, BLUE (RGB) CABLE HANGING FROM THE CEILING (NOT CONNECTED). THE CUSTOMER WAS INSTRUCTED TO SWITCH THE CABLE FROM SDI OUT 1 TO COMP OUT ON THE CV-190, THEN TO TAKE THE CABLE FROM COMP OUT AND CONNECT TO SDI OUT 1 ON THE CV-190, AND THERE WAS NOW AN IMAGE ON THE OEV-191H. THE CUSTOMER WAS THEN INSTRUCTED TO TAKE THE CABLE ON THE OEV-262H AND MOVE IT FROM VIDEO INTO SDI 1 IN, AND IT HAD AN IMAGE. THE ISSUE WAS RESOLVED; THEREFORE, THE DEVICE WILL NOT BE RETURNED TO OLYMPUS FOR EVALUATION. THE INVESTIGATION IS ONGOING AND FOLLOW UP WITH THE USER FACILITY IS CURRENTLY BEING PERFORMED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA III VIDEO SYSTEM CENTER IS NOT DISPLAYING AN IMAGE ON THE OEV-191H / OEV 262H MONITORS. THERE WERE NO REPORTS OF PATIENT HARM.