FDA Adverse Event Malfunction Summary report: N

HAMILTON MEDICAL AG

MDR report key: 17676120 · Received September 4, 2023

Report

Report Number
3001421318-2023-22874
Event Type
Malfunction
Date Received
September 4, 2023
Date of Event
April 23, 2021
Report Date
September 1, 2023
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
00730002856789
PMA / PMN Number
K121225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS PREPARED FOR TREATMENT. THE ROOT CAUSE OF \ "TF 444004 VOLTAGE OUT OF TOLERANCE\ " ALARM WAS DETERMINED TO BE THE DEFECTIVE MAINBOARD (PN 160200). THE CORRECTION IN THIS CASE HAS BEEN THE REMOVAL AND REPLACEMENT OF THE MAINBOARD (PN 160200). AFTER REPLACEMENT THE PROBLEM WAS SOLVED. THERE WAS NO REPORTED HARM TO PATIENT, USER OR THIRD PARTY.

Description of Event or Problem · 0

VENTILATOR TOTALLY IN NOT WORKING CONDITION. MACHINE DIAGNOSED AND FOUND MAINBOARD (160200) FAULTY. CROSS CHECK WITH NEW MAINBOARD (160382).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1492118 HAMILTON MEDICAL AG HAMILTON-C2 CBK HAMILTON MEDICAL AG HAMILTON-C2 00730002856789

Patients

Seq Age Sex Outcome Treatment
1 Unknown