HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-24854
- Event Type
- Malfunction
- Date Received
- September 4, 2023
- Date of Event
- January 18, 2021
- Report Date
- September 1, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K201306
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS NOT IN USE. THE ROOT CAUSE WAS DETERMINED TO BE DEFECTIVE ESM BOARD. IN CONSEQUENCE ESM AND QVENT WERE REPLACED. THERE WAS NO PATIENT OR USER HARM.
UNIT DOESN'T START UP. BOOT PROCESS IS INTERRUPTED. REPLACED ESM BOARD. PN.160565.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1678997 | HAMILTON MEDICAL AG | HAMILTON-C3 | CBK | HAMILTON MEDICAL AG | HAMILTON-C3 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |