FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG SELF TEST 2CT

MDR report key: 17672947 · Received September 1, 2023

Report

Report Number
1221359-2023-01383
Event Type
Malfunction
Date Received
September 1, 2023
Date of Event
July 11, 2023
Report Date
October 12, 2023
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA210264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D4-UDI: (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 218566 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT: 218566, TEST BASE PART NUMBER 195-430H/ LOT: 215334. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Additional Manufacturer Narrative · 0

D4- (B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE USE; DEVICE DISCARDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST FOR TWO (2) TESTS PERFORMED ON MULTIPLE DAYS. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF TWO (2). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023. THE CONSUMER REPORTED BECOMING ILL AND INITIAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA A FLOWFLEX COVID-19 RAPID ANTIGEN TEST GENERATING A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Description of Event or Problem · 0

THE CONSUMER REPORTED FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST FOR TWO (2) TESTS PERFORMED ON MULTIPLE DAYS. THIS MFR. REPORT ADDRESSES TEST TWO (2) OF TWO (2). THE CONSUMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 ANTIGEN SELF-TEST PERFORMED ON (B)(6) 2023. THE CONSUMER REPORTED BECOMING ILL AND INITIAL TESTING WAS PERFORMED ON (B)(6) 2023 VIA A FLOWFLEX COVID-19 RAPID ANTIGEN TEST GENERATING A POSITIVE RESULT. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
858701 BINAXNOW COVID-19 AG SELF TEST 2CT CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 218566 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown