LABSYSTEM PRO
Report
- Report Number
- 2124215-2023-47159
- Event Type
- Malfunction
- Date Received
- September 1, 2023
- Date of Event
- August 9, 2023
- Report Date
- October 2, 2023
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- DRQ
- UDI-DI
- 08714729885948
- PMA / PMN Number
- K152693
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IT WAS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED.
CORRECTION H6 EVALUATION CONCLUSION CODES: CAUSE NOT ESTABLISHED > CAUSE TRACED TO COMPONENT FAILURE. THE DEVICE WAS NOT RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS, HOWEVER, THE LABSYSTEM PRO WAS EVALUATED BY A BOSTON SCIENTIFIC FIELD SERVICE ENGINEER ONSITE. CHECKED CMOS BATTERY FOUND VOLTAGE AT 2.760. REPLACED CMOS BATTERY VOLTAGE ON NEW BATTERY WAS 3.293 V. UPGRADED SOFTWARE TO 4.2. TESTED UNIT, SYSTEM PASSED ALL FUNCTIONAL TESTS AND IS READY FOR USE. ANALYSIS WAS ABLE TO CONFIRM THE REPORTED CLINICAL OBSERVATIONS.
IT WAS REPORTED THAT DURING A PROCEDURE, A LABSYSTEM PRO WAS SELECTED FOR USE. UNIT FROZE DURING THE CASE. ISSUE OCCURRED ABOUT IN THE MIDDLE OF A 3 HOUR CASE. ACCOUNT WAS PACING FROM THE BEGINNING. TROUBLESHOOTING WITH THE ACCOUNT EXTENDED TIMING FOR PACING FROM 45 MINUTES TO 60 MINUTES. PROCEDURE WAS NOT ABLE TO BE COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS OCCURRED. UNIT WILL NOT BE RETURNED SINCE IT WAS REPAIRED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER SEDATION.
THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER SEDATION. IT WAS REPORTED THAT DURING A PROCEDURE, A LABSYSTEM PRO WAS SELECTED FOR USE. UNIT FROZE DURING THE CASE. ISSUE OCCURRED ABOUT IN THE MIDDLE OF A 3 HOUR CASE. ACCOUNT WAS PACING FROM THE BEGINNING. TROUBLESHOOTING WITH THE ACCOUNT EXTENDED TIMING FOR PACING FROM 45 MINUTES TO 60 MINUTES. PROCEDURE WAS NOT ABLE TO BE COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS OCCURRED. UNIT WILL NOT BE RETURNED SINCE IT WAS REPAIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1190266 | LABSYSTEM PRO | AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL | DRQ | BOSTON SCIENTIFIC CORPORATION | 86620 | 1D911712 | 08714729885948 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |