FDA Adverse Event Malfunction Summary report: N

LABSYSTEM PRO

MDR report key: 17670484 · Received September 1, 2023

Report

Report Number
2124215-2023-47159
Event Type
Malfunction
Date Received
September 1, 2023
Date of Event
August 9, 2023
Report Date
October 2, 2023
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
DRQ
UDI-DI
08714729885948
PMA / PMN Number
K152693
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IT WAS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED.

Additional Manufacturer Narrative · 0

CORRECTION H6 EVALUATION CONCLUSION CODES: CAUSE NOT ESTABLISHED > CAUSE TRACED TO COMPONENT FAILURE. THE DEVICE WAS NOT RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS, HOWEVER, THE LABSYSTEM PRO WAS EVALUATED BY A BOSTON SCIENTIFIC FIELD SERVICE ENGINEER ONSITE. CHECKED CMOS BATTERY FOUND VOLTAGE AT 2.760. REPLACED CMOS BATTERY VOLTAGE ON NEW BATTERY WAS 3.293 V. UPGRADED SOFTWARE TO 4.2. TESTED UNIT, SYSTEM PASSED ALL FUNCTIONAL TESTS AND IS READY FOR USE. ANALYSIS WAS ABLE TO CONFIRM THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A PROCEDURE, A LABSYSTEM PRO WAS SELECTED FOR USE. UNIT FROZE DURING THE CASE. ISSUE OCCURRED ABOUT IN THE MIDDLE OF A 3 HOUR CASE. ACCOUNT WAS PACING FROM THE BEGINNING. TROUBLESHOOTING WITH THE ACCOUNT EXTENDED TIMING FOR PACING FROM 45 MINUTES TO 60 MINUTES. PROCEDURE WAS NOT ABLE TO BE COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS OCCURRED. UNIT WILL NOT BE RETURNED SINCE IT WAS REPAIRED. THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER SEDATION.

Description of Event or Problem · 0

THIS EVENT IS BEING REPORTED FOR ABORTED/CANCELLED PROCEDURE WITH A PATIENT UNDER SEDATION. IT WAS REPORTED THAT DURING A PROCEDURE, A LABSYSTEM PRO WAS SELECTED FOR USE. UNIT FROZE DURING THE CASE. ISSUE OCCURRED ABOUT IN THE MIDDLE OF A 3 HOUR CASE. ACCOUNT WAS PACING FROM THE BEGINNING. TROUBLESHOOTING WITH THE ACCOUNT EXTENDED TIMING FOR PACING FROM 45 MINUTES TO 60 MINUTES. PROCEDURE WAS NOT ABLE TO BE COMPLETED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS OCCURRED. UNIT WILL NOT BE RETURNED SINCE IT WAS REPAIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1190266 LABSYSTEM PRO AMPLIFIER AND SIGNAL CONDITIONER, TRANSDUCER SIGNAL DRQ BOSTON SCIENTIFIC CORPORATION 86620 1D911712 08714729885948

Patients

Seq Age Sex Outcome Treatment
1 Unknown