FDA Adverse Event Injury Summary report: N

EDWARDS ESHEATH+ INTRODUCER SET

MDR report key: 17670146 · Received September 1, 2023

Report

Report Number
2015691-2023-15656
Event Type
Injury
Date Received
September 1, 2023
Date of Event
August 10, 2023
Report Date
September 1, 2023
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYB
PMA / PMN Number
K200258
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PER THE INSTRUCTIONS FOR USE (IFU), CARDIOVASCULAR INJURIES SUCH AS PERFORATION OR DISSECTION OF VESSELS, VENTRICLE, MYOCARDIAL OR VALVULAR STRUCTURES ARE KNOWN POTENTIAL RISKS OR ADVERSE EVENTS ASSOCIATED WITH THE OVERALL TRANSCATHETER HEART VALVE (THV) PROCEDURE AND MAY REQUIRE INTERVENTION. ACCORDING TO THE LITERATURE REVIEW, AND AS DOCUMENTED IN A TECHNICAL SUMMARY WRITTEN BY EDWARDS LIFESCIENCES, VASCULAR COMPLICATIONS ARE A WELL-RECOGNIZED COMPLICATION OF THE TRANSFEMORAL THV PROCEDURE IN THIS ELDERLY POPULATION WITH MULTIPLE CO-MORBIDITIES. EDWARDS HAS REVIEWED MANY REPORTS, INCLUDING SCREENING DATA RECORDS AND SOURCE DOCUMENTATION OF VASCULAR COMPLICATIONS, AND HAS FOUND THAT THE ROOT CAUSE IS TYPICALLY RELATED TO A COMBINATION OF VESSEL SIZE, TORTUOSITY, AND CALCIFICATIONS. ALTHOUGH THE INCIDENCE IS DECREASING WITH SMALLER SHEATH/DELIVERY SYSTEM SIZES AND PHYSICIAN EXPERIENCE, THERE WILL CONTINUE TO BE CASES IN WHICH VASCULAR COMPLICATIONS WILL OCCUR. THE THV PHYSICIAN TRAINING MANUALS INSTRUCT ON PROCEDURAL CONSIDERATIONS FOR SHEATH INSERTION WITH REGARDS TO PROPER SCREENING CRITICAL TO REDUCING VASCULAR COMPLICATIONS. THE TRAINING MANUAL INSTRUCTS THE OPERATOR ON PROPER SHEATH INSERTION AND WITHDRAWAL TECHNIQUES, INCLUDING PRE-DILATING THE VESSEL WITH THE EDWARDS DILATORS, AS NEEDED. IT ALSO NOTES THAT CALCIFICATION MAY REDUCE LUMEN DIAMETER AND LIMIT OR PREVENT THE TRANSFEMORAL PASSAGE OF THE DEVICES. THE IFU CONTRAINDICATES PATIENTS WITH ILIO-FEMORAL VESSEL CHARACTERISTICS THAT WOULD PRECLUDE SAFE PLACEMENT OF SHEATHS SUCH AS SEVERE OBSTRUCTIVE CALCIFICATION OR SEVERE TORTUOSITY. PRE-PROCEDURE SCREENING AND ASSESSMENT OF THE FEMORAL/ILIAC ARTERY INTERNAL DIAMETERS WILL ENABLE THE CLINICIAN TO DETERMINE IF THE SAPIEN VALVE CAN BE DELIVERED TRANSFEMORALLY. ASSESSMENT OF LOCATION AND AMOUNT OF CIRCUMFERENTIAL CALCIUM WILL AID IN DETERMINING AREAS OF REDUCED VESSEL DIAMETERS. THE OPERATORS ARE TRAINED TO MEASURE MINIMUM VESSEL DIAMETER TAKING CALCIUM INTO ACCOUNT. THE PHYSICIAN TRAINING MANUAL ALSO LISTS THE MINIMUM RECOMMENDED VESSEL SIZE FOR EACH SIZE DEVICE. DESPITE THE BEST SCREENING TOOLS, A SMALL PERCENTAGE OF PATIENTS WILL HAVE FEMORAL/ILIAC VESSELS THAT ARE NOT AMENABLE TO THE TRANS-FEMORAL APPROACH OR WHERE INCREASED RESISTANCE IS ENCOUNTERED DURING INSERTION OF DEVICES. IN MANY CASES, THE VESSEL MINIMUM LUMINAL DIAMETER (MLD) MAY BE BORDERLINE OR BELOW THE INDICATED SIZE. IN ADDITION, SIGNIFICANT CALCIFICATION AND/OR TORTUOSITY, NOT ALWAYS APPRECIABLE ON IMAGING, COULD BE CONTRIBUTING FACTORS TO THE EVENT. THE MINIMUM REQUIRED VESSEL DIAMETER FOR A 14FR ESHEATH+ IS 5.5 MM. IN THIS CASE, THERE WAS NO ALLEGATION OR INDICATION A DEVICE MALFUNCTION CONTRIBUTED TO THIS ADVERSE EVENT. IN THIS CASE, THERE WAS NO ALLEGATION OR INDICATION A PRODUCT MALFUNCTION CONTRIBUTED TO THIS ADVERSE EVENT. A DEFINITIVE ROOT CAUSE FOR THE DISSECTION COULD NOT BE DETERMINED; HOWEVER, IT MAY BE DUE TO PATIENT FACTORS NOT PROVIDED AND/OR PROCEDURAL FACTORS (NON-EDWARDS CLOSURE DEVICE MALFUNCTION). A REVIEW OF EDWARDS LIFESCIENCES RISK MANAGEMENT DOCUMENTATION WAS PERFORMED FOR THIS CASE. THE REPORTED EVENT IS AN ANTICIPATED RISK OF THE TRANSCATHETER HEART VALVE PROCEDURE, ADDITIONAL ASSESSMENT OF THIS ADVERSE EVENT IS NOT REQUIRED AT THIS TIME. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS MONTHLY, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. AS SUCH, NEITHER A PRODUCT RISK ASSESSMENT, NOR CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME. H3 OTHER TEXT : DEVICE NOT RETURNED.

Description of Event or Problem · 0

AS REPORTED BY OUR EDWARDS LIFESCIENCES JAPANESE AFFILIATE, DURING THE TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE OF A 23 MM SAPIEN 3 ULTRA RESILIA VALVE, UPON REMOVAL OF THE 14FR ESHEATH+, BLEEDING WAS OBSERVED DUE TO A NON-EDWARDS CLOSURE DEVICE MALFUNCTION. HEMOSTASIS WAS ACHIEVED SURGICALLY. A DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) WAS PERFORMED OF THE LOWER EXTREMITIES, AND IT REVEALED A DISSECTION IN THE ILIAC-FEMORAL ARTERY REGION WHICH WAS SURGICALLY REPAIRED. FURTHER ASSESSMENT INDICATED THE CAUSE OF THE DISSECTION MAY HAVE BEEN DAMAGE TO THE INNER MEMBRANE WHEN THE NON-EDWARDS CLOSURE DEVICE WAS APPLIED, BUT IT COULD NOT BE CONFIRMED AS THE DETAILS WERE NOT CLEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1190207 EDWARDS ESHEATH+ INTRODUCER SET INTRODUCER, CATHETER DYB EDWARDS LIFESCIENCES 914ESPJ

Patients

Seq Age Sex Outcome Treatment
1 81 YR Female Required Intervention