FDA Adverse Event Malfunction Summary report: N

INPEN MMT-105NNBLNA NOVO NORDISK BLUE

MDR report key: 17668862 · Received September 1, 2023

Report

Report Number
3012822846-2023-01549
Event Type
Malfunction
Date Received
September 1, 2023
Date of Event
August 15, 2023
Report Date
November 9, 2023
Manufacturer
COMPANION MEDICAL INC
Product Code
FMF
UDI-DI
000010862088000344
PMA / PMN Number
K201337
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER REPORT: DOSE DIAL IS DIFFICULT TO TURN/DIAL. PER VISUAL INSPECTION: NO PHYSICAL DAMAGE TO CARTRIDGE HOLDER OR INPEN FRONT AND BACK SHELL WAS NOTED. CARTRIDGE HOLDER LOCKS PROPERLY IN PLACE. INPEN PAIRED WITH COMMERCIAL MOBILE APP WITH SMALL DOSE. PERFORMED ENCODER BOND INVESTIGATION AND FOUND THAT THE ENCODER PATTERN WHEEL TABS ROTATING AND TRAVELING OFF THE KEYED SLOTS OF DOSE NUT GUIDES. ENCODER PATTERN WHEEL SHOULD NEVER ROTATE. THIS CAUSES AN UNEXPECTED TRAVEL OF THE ENCODER PATTERN WHEEL CREATING RESISTANCE TO DIAL AND DISPENSING. ALSO, PLASTIC SHAVINGS CONTAMINATION BUILDS UP FOUND CAUSED BY ENCODER WHEEL TABS RUBBING AT THE WALLS OF THE DOSE NUT. ABRASIONS DOWN THE LENGTH OF THE DOSE NUT. INPEN PASSED AND IS WITHIN SPECIFICATION (CW: 0.75 OZF-IN AND CCW: 5.25 OZF-IN). UNABLE TO PERFORM BASELINE/WIRELESS FUNCTIONALITY AND DISPLACEMENT DOSE ACCURACY TEST. INPEN PASSED FRONT CAP INVESTIGATION. IN CONCLUSION: IT WAS DETERMINED FROM DESTRUCTIVE ANALYSIS THAT THE DIFFICULT TO DIAL/DOSE WAS CAUSED BY A PATTERN WHEEL MISALIGNMENT DUE TO AN ENCODER BASE BOND FAILURE. THIS CAN AFFECT INSULIN DELIVERY. THEREFORE, THE CUSTOMER CONCERN OF DOSE DIAL BEING DIFFICULT TO DIAL/DOSE WAS CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED DOSE KNOB WAS DIFFICULT TO TURN. THE TROUBLESHOOTING WAS PERFORMED AND THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE. THE DEVICE WILL BE RETURNED FOR PRODUCT ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1262370 INPEN MMT-105NNBLNA NOVO NORDISK BLUE SYRINGE, PISTON FMF COMPANION MEDICAL INC MMT-105NNBLNA B0947 000010862088000344

Patients

Seq Age Sex Outcome Treatment
1 76 YR Male