FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 17666206 · Received August 31, 2023

Report

Report Number
1000113657-2023-00449
Event Type
Malfunction
Date Received
August 31, 2023
Date of Event
August 8, 2023
Report Date
August 31, 2023
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: 141940-1. METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIPS TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-020: USER'S TEST STRIP HAD POOR STORAGE. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 18-AUG-2023 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HI AND HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF HI, 589 AND 554 MG/DL. THE CUSTOMER DOES NOT KNOW HER EXPECTED BLOOD GLUCOSE TEST RESULT RANGE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION (KITCHEN). THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 12/31/2024 AND OPEN VIAL DATE WAS NOT DISCLOSED. THE CUSTOMER DID NOT HAVE ANOTHER VIAL OF TEST STRIPS THAT HAD BEEN STORED AND HANDLED CORRECTLY. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: HI. DATE:(B)(6) 2023. TIME: 9:14AM NON-FASTING. RESULT 2: HI. DATE: (B)(6) 2023. TIME: 7:33AM FASTING. RESULT 3: HI. DATE: (B)(6) 2023. TIME: 6:08 FASTING. RESULT 4: 589 MG/DL. DATE: (B)(6) 2023. TIME: 7:22PM FASTING. RESULT 5: 554 MG/DL. DATE: (B)(6) 2023. TIME: 7:06PM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1437816 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, LEADER TMX 50CTMM MG/DL ZB5293S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 Unknown