FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM
MDR report key: 1766530
·
Received July 22, 2010
Report
- Report Number
- 2122870-2010-00353
- Event Type
- Malfunction
- Date Received
- July 22, 2010
- Date of Event
- May 25, 2010
- Report Date
- July 21, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC WAS WITHIN SPECIFICATIONS PRIOR TO THE EVENT. THE CUSTOMER SENT SAMPLE TO CUSTOMER PRODUCT LINE SUPPORT (CPLS) FOR ADDITIONAL TESTING. CPLS TESTING CONFIRMS ROOT CAUSE TO BE BIOTIN INTERFERENCE (PATIENT SOURCE INTERFERENCE) IN THE TT3 ASSAY. SERVICE WAS NOT DISPATCHED FOR THIS EVENT.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING ABOVE THE NORMAL REFERENCE RANGE RESULTS FOR TOTAL TRIIODOTHYRONINE (TT3) AND THYROID-STIMULATING HORMONE (TSH) GENERATED BY THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM FOR ONE PATIENT. SUBSEQUENT TESTING ON AN ALTERNATE METHODOLOGY PRODUCED LOWER RESULTS FOR BOTH ANALYTES. THE CUSTOMER DID NOT REPORT AFFECT TO THE PATIENT OR USER ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |