FDA Adverse Event Injury Summary report: N

LINEAR ST

MDR report key: 17662067 · Received August 31, 2023

Report

Report Number
3006630150-2023-05210
Event Type
Injury
Date Received
August 31, 2023
Date of Event
April 25, 2023
Report Date
August 31, 2023
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
UDI-DI
08714729767749
PMA / PMN Number
P030017
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS, UPN: M365SC2218700, MODEL: SC-2218-70, SERIAL: (B)(6), BATCH: 7082672.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT WAS NOT GETTING AN ADEQUATE PAIN RELIEF FROM THE SPINAL CORD STIMULATOR (SCS) DEVICE. IT WAS NOTED THAT THE PATIENTS RIGHT LEAD HAD MIGRATED WHICH WAS CONFIRMED VIA X-RAY. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE LINEAR LEADS WERE REPLACED WITH A PADDLE LEAD. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WERE DISCARDED BY THE MEDICAL FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1602485 LINEAR ST STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW BOSTON SCIENTIFIC NEUROMODULATION SC-2218-70 7082625 08714729767749

Patients

Seq Age Sex Outcome Treatment
1 59 YR Male Required Intervention