FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRIC INFUSION PUMP SINGLE CHANNEL RECERT

MDR report key: 1765757 · Received July 21, 2010

Report

Report Number
6000001-2010-01854
Event Type
Malfunction
Date Received
July 21, 2010
Date of Event
February 19, 2010
Report Date
July 12, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DURING A REVIEW OF THE EVENT HISTORY, THE OCCURRENCE DATE IS (B)(6) 2010. THERE IS A CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION ASSOCIATED WITH THIS REPORT, MDQ-CAPA-(B)(4).

Description of Event or Problem · 1

DURING SERVICE AT BAXTER, A TECHNICIAN REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11THAT WAS CAUSED BY THE AIL PCB (AIR IN LINE PRINTED CIRCUIT BOARD) THAT WAS OUT OF CALIBRATION AND WAS RECALIBRATED BY SERVICE. THE FACILITY REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY, ADVERSE EVENT OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE (B)(4) CATEGORIZED AS REMEDIATED.

Description of Event or Problem · 1

DURING A PATIENT CHOLECYSTECTOMY PROCEDURE, THE SURGEON INSERTED, THROUGH THE SAME INCISION CUT AS THE SPIDER DEVICE, A RIGID SUCTION/IRRIGATION INSTRUMENT. THE SURGEON ACCIDENTALLY TANGLED THE SUCTION/IRRIGATION INSTRUMENT INTO THE SPIDER DEVICE AND BROKE OR SNAPPED LOOSE THE UPPER RIGHT LINK ARM AT THE LINK ARM CONNECTOR. A PIECE OF THE LINK ARM CONNECTOR FELL INTO THE SURGICAL BED. THE SURGEON WAS ABLE TO LOCATE AND RETRIEVE THE PIECE FROM THE PATIENT. NO INJURY OR IMPACT TO PATIENT CARE WAS REPORTED. THE DEVICE WAS RETURNED TO TRANSENTERIX, INC. FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRIC INFUSION PUMP SINGLE CHANNEL RECERT PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1