HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-12061
- Event Type
- Malfunction
- Date Received
- August 30, 2023
- Date of Event
- August 2, 2022
- Report Date
- August 29, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- PMA / PMN Number
- K201658
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE DEVICE FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS PREPARED FOR TREATMENT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE PRESSURE SENSOR ASSEMBLY (PSA). IN CONSEQUENCE THE PRESSURE SENSOR ASSEMBLY HAS BEEN REPLACED. THERE WAS NO PATIENT OR USER HARM.
VOLUME- AND FLOW-CURVES ARE UNSTABLE AND DOES NOT GIVE THE RIGHT WAVES ON THE TEST LUNG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1287609 | HAMILTON MEDICAL AG | HAMILTON-C6 | CBK | HAMILTON MEDICAL AG | HAMILTON-C6 | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |