FDA Adverse Event Injury Summary report: N

HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE

MDR report key: 1764484 · Received July 16, 2010

Report

Report Number
2024168-2010-01411
Event Type
Injury
Date Received
July 16, 2010
Date of Event
June 18, 2010
Report Date
June 21, 2010
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
DQX
PMA / PMN Number
K013833
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HI-TORQUE BMW (PART# 1009660, LOT 0032475), INDICATED IS BEING FILED UNDER A SEPARATE MFR #. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS REPORTED. A REVIEW OF DEVICE HISTORY RECORD IS FORTHCOMING. A F/U WILL BE SUBMITTED WITH ALL RELEVANT INFO.

Description of Event or Problem · 1

DEVICE ISSUE: SEPARATED GUIDE WIRE TIP. TIME OF DEVICE ISSUE: DURING THE PROCEDURE. ADVERSE EVENT: TIP COMPRESSED TO VESSEL WALL WITH A STENT. ONSET OF ADVERSE EVENT: DURING THE PROCEDURE. IT WAS REPORTED THAT THE BALANCE MIDDLEWEIGHT (BMW) WAS PLACED IN THE HEAVILY CALCIFIED CIRCUMFLEX (CX) BRANCH OF THE LEFT CORONARY ARTERY AND AFTER THE STENTING, DURING AN ATTEMPT TO REMOVE THE GUIDE WIRE, THERE WAS RESISTANCE AND THE GUIDE WIRE SEPARATED. THIS OCCURRED WITH TWO BMW GUIDE WIRES IN ONE PT. ONE OF THE SEPARATED GUIDE WIRE TIPS WAS COMPRESSING AGAINST THE VESSEL WALL WITH A NON-ABBOTT STENT AND THE OTHER WAS LEFT IN THE PT'S ANATOMY UNTREATED. THE PT IS DOING WELL AND WILL RECEIVE ANTICOAGULATION MEDICATION (3 AGENTS) FOR THE DURATION OF 12 MONTHS AND ON THE 13TH MONTH, THE PT WILL RECEIVE ONLY TWO AGENTS. NO ADDITIONAL EVENT OR PT INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE DQX ABBOTT VASCULAR-CARDIAC THERAPIES NA 0032475

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention| S GUIDE WIRE: HI-TORQUE BMW| (PART 1009660, LOT 0032475)