FDA Adverse Event Malfunction Summary report: N

CLYDESDALE PTC SPINAL SYSTEM

MDR report key: 17629479 · Received August 28, 2023

Report

Report Number
1030489-2023-00612
Event Type
Malfunction
Date Received
August 28, 2023
Date of Event
August 3, 2023
Report Date
December 6, 2023
Manufacturer
MSD DEGGENDORF MFG
Product Code
MAX
UDI-DI
00643169191075
PMA / PMN Number
K133205
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PRODUCT # 4922055, LOT # 04LA: VISUAL AND OPTICAL INSPECTION CONFIRMED ABOUT CAGE HAS BROKEN AT THE INSERTER ATTACHMENT AREA. OPTICAL INSPECTION OF THE FRACTURE SURFACE DID NOT REVEAL ANY DAMAGE THAT COULD CONTRIBUTE TO THE IMPLANT BREAKING. SURFACE SCRATCHES ARE NOTED ON THE TIP/NOSE AND SOME ON THE RIBS.THE DISC SPACE WAS A LITTLE TOO NARROW FOR THE SPACER AS STATED IN THE COMPLAINT DESCRIPTION. IT APPEARS THE EXCESSIVE FORCE DAMAGED THE IMPLANT DURING THE INSERTION PROCESS. PRODUCT# 4986050, LOT# 70MW: VISUAL AND OPTICAL INSPECTION CONFIRMED ABOUT CAGE HAS BROKEN AT THE INSERTER ATTACHMENT AREA. OPTICAL INSPECTION OF THE FRACTURE SURFACE DID NOT REVEAL ANY DAMAGE THAT COULD CONTRIBUTE TO THE IMPLANT BREAKING. SURFACE SCRATCHES ARE NOTED ON THE TIP/NOSE AND SOME ON THE RIBS.THE DISC SPACE WAS A LITTLE TOO NARROW FOR THE SPACER AS STATED IN THE COMPLAINT DESCRIPTION. IT APPEARS THE EXCESSIVE FORCE DAMAGED THE IMPLANT DURING THE INSERTION PROCESS. PRODUCT# 4986055, LOT# 52NE: VISUAL AND OPTICAL INSPECTION CONFIRMED ABOUT CAGE HAS BROKEN AT THE INSERTER ATTACHMENT AREA. OPTICAL INSPECTION OF THE FRACTURE SURFACE DID NOT REVEAL ANY DAMAGE THAT COULD CONTRIBUTE TO THE IMPLANT BREAKING. SURFACE SCRATCHES ARE NOTED ON THE TIP/NOSE AND SOME ON THE RIBS.THE DISC SPACE WAS A LITTLE TOO NARROW FOR THE SPACER AS STATED IN THE COMPLAINT DESCRIPTION. IT APPEARS THE EXCESSIVE FORCE DAMAGED THE IMPLANT DURING THE INSERTION PROCESS. PRODUCT# 4986055, LOT# 85MN: VISUAL AND OPTICAL INSPECTION CONFIRMED ABOUT CAGE HAS BROKEN AT THE INSERTER ATTACHMENT AREA. OPTICAL INSPECTION OF THE FRACTURE SURFACE DID NOT REVEAL ANY DAMAGE THAT COULD CONTRIBUTE TO THE IMPLANT BREAKING. SURFACE SCRATCHES ARE NOTED ON THE TIP/NOSE AND SOME ON THE RIBS.THE DISC SPACE WAS A LITTLE TOO NARROW FOR THE SPACER AS STATED IN THE COMPLAINT DESCRIPTION. IT APPEARS THE EXCESSIVE FORCE DAMAGED THE IMPLANT DURING THE INSERTION PROCESS. PRODUCT# 4922055, LOT# 10FL: VISUAL AND OPTICAL INSPECTION CONFIRMED ABOUT CAGE HAS BROKEN AT THE INSERTER ATTACHMENT AREA. OPTICAL INSPECTION OF THE FRACTURE SURFACE DID NOT REVEAL ANY DAMAGE THAT COULD CONTRIBUTE TO THE IMPLANT BREAKING. SURFACE SCRATCHES ARE NOTED ON THE TIP/NOSE AND SOME ON THE RIBS.THE DISC SPACE WAS A LITTLE TOO NARROW FOR THE SPACER AS STATED IN THE COMPLAINT DESCRIPTION. IT APPEARS THE EXCESSIVE FORCE DAMAGED THE IMPLANT DURING THE INSERTION PROCESS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 4986050, SERIAL/LOT #: (B)(6), UBD: 29-AUG-2030, UDI#: (B)(4); PRODUCT ID: 4986055, SERIAL/LOT #: (B)(6), UBD: 14-NOV-2030, UDI#: (B)(4); PRODUCT ID: 4986055, SERIAL/LOT #: (B)(6), UBD: 13-JUN-2030, UDI#: (B)(4); PRODUCT ID: 4922055, SERIAL/LOT #: (B)(6), UBD: 05-JUN-2026, UDI#: (B)(4). MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING OLIF25 3 VERTEB RAE (L2/3, L3/4, L4/5) FOR LUMBAR DEGENERATIVE SCOLIOSIS. IT WAS REPORTED THAT THE DISC HEIGHT WAS NARROW (4 - 5 MM), BUT THE DOCTOR WANTED TO INSERT A CAGE WITH A GREATER HEIGHT. AFTER THE TRIAL, THE INSERTION OF THE PRODUCT WAS TRIED, BUT IT CRACKED. FIRST, A CAGE WAS PLACED IN L4/5. NEXT, 2 CAGES BROKE DURING CAGE INSERTION IN L3/4, AND FINALLY A  CAGE WAS PLACED. THREE PIECES BROKE DURING CAGE INSERTION IN THE FINAL L2/3, AND FINALLY ANOTHER CAGE WAS PLACED. THE BROKEN CAGES AND BROKEN FRAGMENTS WERE SECURED, BUT IT IS UNKNOWN IF THE BROKEN FRAGMENTS WERE LEFT BEHIND IN THE BODY BECAUSE THEY COULD NOT BE FLUOROSCOPICALLY CHECKED. THERE WAS A DELAY IN THE PROCEDURE BY MORE THAN 60 MINUTES. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1338753 CLYDESDALE PTC SPINAL SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, MAX MSD DEGGENDORF MFG 4922055 04LA 00643169191075

Patients

Seq Age Sex Outcome Treatment
1 Unknown