AEON ENDOSCOPIC STAPLER
Report
- Report Number
- 3012998149-2023-00004
- Event Type
- Injury
- Date Received
- August 25, 2023
- Date of Event
- July 25, 2023
- Report Date
- August 25, 2023
- Manufacturer
- LEXINGTON MEDICAL INC.
- Product Code
- GDW
- UDI-DI
- 00851574008943
- PMA / PMN Number
- 222210
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AFTER INVESTIGATION, THE LEVEL ONE ROOT CAUSE FOR BLEEDING WAS THAT THE SURGEON HAD FIRED THE RELOAD WITH A HEM-O-LOCK IN THE RELOAD JAWS, RESULTING IN MULTIPLE MALFORMED STAPLES. FIRING A RELOAD OVER AN OBSTRUCTION IS KNOWN TO RESULT IN STAPLE MALFORMATION AND THEREFORE POOR HEMOSTATIC PERFORMANCE. THE IFU USED FOR THIS REGION IS MOST LIKELY THE SPANISH LANGUAGE VERSION OF 900298. IN THE SPANISH VERSION OF 900298 REV-T (THE MOST RECENT REVISION THAT WAS APPROVED BY COLOMBIAN AUTHORITIES AT TIME OF THE EVENT), WARNING #5 ON PAGE 7 INDICATES THAT FIRING OVER AN OBSTRUCTION CAN RESULT IN INCORRECTLY FORMED STAPLES. THE LEVEL TWO ROOT CAUSE IS THEREFORE USER ERROR.
FIVE 45 GREY CURVED RELOADS AND ONE 30 WHITE CURVED RELOAD WERE USED FOR A LOBECTOMY PROCEDURE THAT WAS PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE THE SURGEON FIRED THE RELOAD TO SEAL AN ARTERY, THE STAPLE LINE DID NOT FULLY SEAL THE ARTERY AND THE PATIENT BLED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1803237 | AEON ENDOSCOPIC STAPLER | AEON ENDOSCOPIC STAPLER | GDW | LEXINGTON MEDICAL INC. | AESC45G | 22BR541 | 00851574008943 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention | AESC45G AEON ENDOSCOPIC STAPLER 2.0MM/45MM CURVED| AESC45G AEON ENDOSCOPIC STAPLER 2.0MM/45MM CURVED| AESC45G AEON ENDOSCOPIC STAPLER 2.0MM/45MM CURVED| AESC45G AEON ENDOSCOPIC STAPLER 2.0MM/45MM CURVED| ASR30WC AEON ENDOSCOPIC STAPLER 2.5MM/30MM CURVED |