Z.S.G.M. CUTTER 1.5:1 RATIO ( CAUTION: SHARP )
Report
- Report Number
- 0001526350-2023-00978
- Event Type
- Malfunction
- Date Received
- August 25, 2023
- Date of Event
- March 3, 2023
- Report Date
- August 25, 2023
- Manufacturer
- ZIMMER SURGICAL, INC.
- Product Code
- FZW
- PMA / PMN Number
- PREAMENDMENT
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS EVENT HAS BEEN RECORDED BY ZIMMER BIOMET UNDER (B)(4). G2 FOREIGN: (B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: 0001526350-2023-00309-1. THE INVESTIGATION IS COMPLETE. THIS IS AN INITIAL FINAL REPORT SUBMISSION. REVIEW OF THE MOST RECENT REPAIR RECORD IDENTIFIED THE FOLLOWING RELATED FINDING: TECH FOUND THE CUTTER DID NOT PASS THE SAMPLE MESH TEST. THE CUSTOMER WAS NOTIFIED OF THE FAILURE AS REPAIRS ARE NOT PERFORMED FOR ZGSM ACCESSORIES. REVIEW OF THE DEVICE HISTORY RECORD IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. THE EVENT IS CONFIRMED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THE UNIT IS NOT RATCHETING. NO HARM OR DELAY REPORTED. NO ADVERSE EVENTS WERE REPORTED AS A RESULT OF THIS MALFUNCTION. DUE DILIGENCE IS COMPLETE. AT PRODUCT EVALUATION INVESTIGATION THE CUTTER DID NOT PASS THE SAMPLE MESH TEST. NO ADDITIONAL EVENT INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2077224 | Z.S.G.M. CUTTER 1.5:1 RATIO ( CAUTION: SHARP ) | EXPANDER, SURGICAL, SKIN GRAFT | FZW | ZIMMER SURGICAL, INC. | N/A | 33327000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Prefer Not To Disclose | CUTTER 3:1 RATIOPN 00770303000LN UNKNOWN |