630G INSULIN PUMP MMT-1715K 630G BLACK MG
Report
- Report Number
- 2032227-2023-261524
- Event Type
- Malfunction
- Date Received
- August 25, 2023
- Date of Event
- August 9, 2023
- Report Date
- October 12, 2023
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- UDI-DI
- 000000763000316631
- PMA / PMN Number
- P150001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- 003
Narratives
THE PUMP WAS RECEIVED WITH A PARTIALLY BROKEN BATTERY TUBE THREADS. THE PUMP PASSED THE SELF TEST, DISPLACEMENT TEST AND ACTIVE CURRENT MEASUREMENT. HOWEVER, THE PUMP HAD A HIGH SLEEP CURRENT MEASUREMENT. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS OR ALERTS NOTED. NO POWER ERROR 25 AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM RECORDED IN THE FORMATTED HISTORY FILE ON THE EVENT DATE. HOWEVER, INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/08/2023 23:05:42.000, 08/08/2023 23:15:00.000, 08/09/2023 14:21:04.000, 08/09/2023 14:31:00.000, 08/09/2023 17:21:08.000, 08/09/2023 17:21:08.000, 08/09/2023 17:23:22.000, 08/09/2023 17:33:00.000, 08/09/2023 18:09:54.000,08/09/2023 18:19:00.000, 08/09/2023 19:00:17.000, 08/09/2023 19:10:00.000, 08/09/2023 19:21:10.000, 08/09/2023 19:28:33.000, 08/09/2023 19:37:43.000, 08/09/2023 19:40:22.000, 08/09/2023 19:41:02.000, 08/09/2023 19:41:47.000, 08/09/2023 19:42:30.000, 08/09/2023 19:45:12.000, 08/09/2023 19:45:28.000 AND 08/09/2023 19:46:10.000. LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/09/2023 19:41:00.000, POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/08/2023 23:21:54.000, 08/09/2023 19:22:22.000, 08/09/2023 19:53:40.000, 08/09/2023 19:53:51.000. PUMP ERROR 23 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/09/2023 19:53:03.000. FAILED BATTERY ALERT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/09/2023 17:21:19.000, 08/09/2023 17:23:03.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND SLIGHT CORROSION ON THE PCBA 1 AND PCBA 2. NO CORROSION OR MOISTURE DAMAGE FOUND ON THE FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. THE ORIGINAL PCBA 1 WAS CLEANED, INSTALLED IN A TEST PCBA 2, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP FAILED THE SLEEP CURRENT MEASUREMENT. THE ORIGINAL PCBA 2 WAS CLEANED, INSTALLED IN A TEST PCBA 1, CASE, INTERNAL BATTERY AND MOTOR. POWERED THE PUMP ON USING THE BATTERY SIMULATOR AND CONTINUED TESTING. THE PUMP FAILED THE SLEEP CURRENT MEASUREMENT. THE PUMP HAD A HIGH SLEEP CURRENT MEASUREMENT (UNEXPECTED BATTERY POWER LOSS) DUE TO SLIGHT CORROSION ON THE PCBA 1 AND PCBA 2. PLEASE SEE BELOW FOR PUMP ERRORS/ALARMS NOTED 1 WEEK PRIOR TO THE EVENT DATE 09-AUG-2023 IN THE FORMATTED HISTORY FILE. PUMP ERROR 53 ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: 08/09/2023 17:21:39.000, 08/09/2023 17:23:22.000. PUMP ERROR 53 ALARM DUE TO SOFTWARE ERROR (LINENUMBER = 5632 ; FILENUMBER = 2005). TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY, A KEYPAD OVERLAY TEXTURE DAMAGE, A SCRATCHED CASE, A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT AND A SERIAL NUMBER LABEL FADING. COSMETIC DAMAGE WAS CONFIRMED AT THE TOP OF BATTERY COMPARTMENT. A HIGH SLEEP CURRENT MEASUREMENT (UNEXPECTED BATTERY POWER LOSS) WAS CONFIRMED DUE TO SLIGHT CORROSION ON THE PCBA 1 AND PCBA 2. PUMP ERROR 53 ALARM (LINENUMBER = 5632 ; FILENUMBER = 2005) WAS CONFIRMED ACCORDING IN THE FORMATTED HISTORY FILE DUE TO SOFTWARE ERROR MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER HAS REPORTED A BATTERY NOT WORKING CORRECTLY AND A CRACK ON THE TOP OF THE BATTERY COMPARTMENT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED AND FOUND THAT THE ISSUE WAS NOT RESOLVED. THE CUSTOMER WILL DISCONTINUE TO USE THE DEVICE AND THE PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1815436 | 630G INSULIN PUMP MMT-1715K 630G BLACK MG | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1715K | HG400MM | 000000763000316631 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Male |