VOALTE NURSE CALL
Report
- Report Number
- 2027454-2023-00067
- Event Type
- Malfunction
- Date Received
- August 24, 2023
- Date of Event
- August 2, 2023
- Report Date
- August 24, 2023
- Manufacturer
- HILL-ROM CARY
- Product Code
- ILQ
- UDI-DI
- 00887761995079
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE HILLROM TECHNICIAN FOUND THE CODE BLUE WAS NOT CONFIGURED FOR SOME ROOMS. THE VOALTE® PATIENT SAFETY APPLICATION IS INTENDED TO COLLECT MEDICAL DEVICE DATA, INCLUDING THERAPY SURFACE AND PATIENT PARAMETERS, TO DOCUMENT ACTIVITIES AND TO ALERT CAREGIVERS AND OTHER PERSONNEL BASED ON CAREGIVER SPECIFIED CRITERIA AND SYSTEM EVENTS. THESE ALERTS AND NOTIFICATIONS WILL BE SENT VIA NETWORK APPLICATION, NURSE COMMUNICATION (OPTIONAL), AND WIRELESS COMMUNICATION DEVICES WITHIN THE HEALTHCARE ENVIRONMENT. THE APPLICATION WILL PROVIDE REPORTS ON THE DATA COLLECTED AND ALERTS GENERATED. THE APPLICATION WILL ALSO PROVIDE DEFAULT VOALTE® PATIENT SAFETY PROTOCOLS WHICH ALLOW CAREGIVERS TO CUSTOMIZE VOALTE®PATIENT SAFETY PROTOCOLS, WITHIN THE APPLICATION, BASED ON THEIR RISK ASSESSMENT OF THE PATIENT. PATIENT SAFETY WILL MONITOR THE VOALTE® PATIENT SAFETY PROTOCOLS SET BY THE CAREGIVER AND THE ASSOCIATED DATA POINTS AND ALERT CAREGIVERS WHEN THE PROTOCOL SETTINGS ARE VIOLATED OR REQUIRE ATTENTION. THE APPLICATION INCLUDES WORKFLOW CAPABILITIES THAT ALLOW TURNING OFF BED EXIT AND SUPPRESSION (OR SILENCING) OF ASSOCIATED ALERTS AS WELL AS AUTOMATIC ENABLING OF BED EXIT AND ASSOCIATED ALERTS. THE HRC TECHNICIAN CORRECTED THE ISSUE IN ECT FOR THE CUSTOMER TO RESOLVE THE ISSUE. ALTHOUGH THERE WAS NO INJURY ASSOCIATED WITH THIS EVENT, IF THE REPORTED ISSUE WERE TO RECUR IT COULD LEAD TO INJURY OR DEATH. THEREFORE HILLROM IS REPORTING THIS COMPLAINT.
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE CODE BLUE IS NOT AUDIBLE AT THE OPERATING ROOMS, THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1038177 | VOALTE NURSE CALL | SYSTEM, COMMUNICATION, POWERED | ILQ | HILL-ROM CARY | 00887761995079 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |