ACCU-CHEK ® COMFORT CURVE TEST STRIPS

MDR report key: 1761345 · Received July 19, 2010

Report

Report Number
1823260-2010-04200
Event Type
Malfunction
Date Received
July 19, 2010
Date of Event
June 30, 2010
Report Date
August 16, 2010
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
K010362
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE CALL ENDED BEFORE THE CUSTOMER PROVIDED ANY FURTHER INFORMATION.

Description of Event or Problem · 1

THE USER EXPERIENCED ERRATIC CO2 RESULTS FOR MULTIPLE PATIENTS, THE EXACT NUMBER IS UNKNOWN. HE PROVIDED CO2 RESULTS FOR TWO PATIENT SAMPLES: PATIENT 1, INITIAL CO2 RESULT WAS 43.9 MMOL/L. THE SAMPLE REPEATED ON THE SAME ANALYTICAL D MODULE ((B)(4)) GAVE 28.7 MMOL/L. THE REPEAT RESULT WAS DEEMED TO BE CORRECT AND WAS REPORTED OUTSIDE THE LABORATORY. PATIENT 2, INITIAL CO2 RESULT WAS ZERO. THE SAMPLE REPEATED ON THE SAME ANALYTICAL D MODULE ((B)(4)) GAVE A NORMAL RESULT, THE EXACT VALUE WAS NOT PROVIDED. THE REPEAT RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE PATIENTS WERE NOT AFFECTED BY THE ISSUE. THE CO2 REAGENT LOT NUMBER WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE SAMPLE PROBES WERE MISALIGNED AND WERE HITTING THE RIGHT SIDE OF THE OUTER CUVETTES. HE CHANGED CUVETTES AND PERFORMED A PROBE ADJUSTMENT. THE FIELD SERVICE REPRESENTATIVE ALSO CLEANED THE VACUUM LINES AND VERIFIED ALL RINSE STATION ADJUSTMENTS WERE PROPERLY ALIGNED. THE USER SUCCESSFULLY RAN CO2 CALIBRATION AND QUALITY CONTROL.

Description of Event or Problem · 1

REPORTER ALLEGED OBTAINING A RESULT OF 217 MG/DL ON THE ADVANTAGE SYSTEM COMPARED BACK TO BACK WITH A RESULT OF 111 MG/DL ON THE DOCTOR'S SYSTEM WHEN TESTING WAS PERFORMED WITHIN 10 MINUTES. NO ACTIONS WERE REPORTED TAKEN OR TREATMENT RECEIVED. NO ADVERSE EVENT REPORTED. A REQUEST WAS MADE FOR THE RETURN OF THE AFFECTED PRODUCT.