FDA Adverse Event Malfunction Summary report: N

TERUMO CDI 500 BLOOD PARAMETER MONITOR

MDR report key: 1761078 · Received July 14, 2010

Report

Report Number
1828100-2010-00841
Event Type
Malfunction
Date Received
July 14, 2010
Date of Event
June 25, 2010
Report Date
July 14, 2010
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DRY
PMA / PMN Number
K972962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

EVAL IN PROGRESS, BUT NOT YET CONCLUDED.

Description of Event or Problem · 1

DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THE SERVICE TECH REPORTED THAT THE PROBE FAILED THE 3000 VOLT HYPOT TEST. THE MONITOR WAS ORIGINALLY RETURNED FOR REPAIR OF A SCREEN FAILURE WHEN THE PROBE FAILURE WAS FOUND. SINCE THIS EVENT OCCURRED DURING ROUTINE TESTING, THERE WAS NO PATIENT INVOLVEMENT DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TERUMO CDI 500 BLOOD PARAMETER MONITOR ON LINE BLOOD GAS MONITOR DRY TERUMO CARDIOVASCULAR SYSTEMS CORP. 500AHCT

Patients

Seq Age Sex Outcome Treatment
1