FDA Adverse Event Malfunction Summary report: N

HEARTMATE 3 VAD MODULAR CABLE

MDR report key: 17610436 · Received August 24, 2023

Report

Report Number
2916596-2023-05923
Event Type
Malfunction
Date Received
August 24, 2023
Date of Event
August 4, 2023
Report Date
August 24, 2023
Manufacturer
THORATEC CORPORATION
Product Code
DSQ
UDI-DI
00813024013181
PMA / PMN Number
P160054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION CONCLUSION: THE REPORTED EVENT OF DRIVELINE POWER FAULT ALARMS WAS CONFIRMED VIA ANALYSIS OF THE LOG FILES; HOWEVER, THE REPORTED ALARMS WERE NOT REPRODUCED DURING TESTING OF THE RETURNED OF THE MODULAR CABLE (LOT NUMBER: 8140086). THE REPORTED EVENT OF KINKS BEING OBSERVED ON THE MODULAR CABLE WAS CONFIRMED VIA EVALUATION OF THE RETURNED MODULAR CABLE. THE CONTROLLER EVENT LOG FILES CONTAINED APPROXIMATELY 1.5 HOURS OF DATA COMBINED ((B)(6) 2023 FROM 14:12:52 TO 15:29:14 PER THE TIMESTAMP). A DRIVELINE POWER FAULT ALARM ASSOCIATED WITH A POWER A BROKEN FAULT WAS OBSERVED THROUGHOUT THE LOG FILE DUE TO THE CURRENT SENSE ON LINE A DROPPING BELOW THE THRESHOLD AMPERAGE. THE DRIVELINE WAS DISCONNECTED ON (B)(6) 2023 AT 15:43:20 TO EXCHANGE THE CONTROLLER. THE INITIAL CONDITIONS THAT CAUSED THE DRIVELINE POWER FAULT ALARM TO ACTIVATE WERE NOT CAPTURED IN THE LOG FILE AND THE ALARM WAS NOT OBSERVED TO RESOLVE WHILE THE CONTROLLER WAS IN USE. THERE WERE NO OTHER NOTABLE ALARMS THROUGHOUT THE LOG FILES. THE PUMP MAINTAINED A SPEED ABOVE THE LOW SPEED LIMIT WHILE THE DRIVELINE WAS CONNECTED. VISUAL INSPECTION OF THE RETURNED MODULAR CABLE REVEALED MINOR KINKS ON THE POLYURETHANE AT APPROXIMATELY 5 INCHES FROM THE INLINE CONNECTOR AND AT APPROXIMATELY 9 INCHES FROM THE CONTROLLER CONNECTOR. THE RETURNED MODULAR CABLE WAS CONNECTED TO THE RETURNED SYSTEM CONTROLLER (SERIAL NUMBER: HSC-104870) AND A MOCK CIRCULATORY LOOP WITHOUT ISSUE. THE MOCK CIRCULATORY LOOP OPERATED ABOVE THE LOW SPEED LIMIT WITHOUT ANY ISSUES OR ATYPICAL ALARMS PRODUCED, INCLUDING WHEN THE MODULAR CABLE WAS MANIPULATED BY HAND. THE INTERNAL WIRES INTEGRITY WAS TESTED AND NO ISSUES WERE OBSERVED. THE POLYURETHANE JACKET AND UNDERLYING LAYERS WERE REMOVED TO INSPECT THE UNDERLYING WIRES AND NO ISSUES WERE OBSERVED. IT WAS REPORTED THAT NO FURTHER ALARMS HAVE BEEN OBSERVED AFTER EXCHANGING THE CONTROLLER AND THE MODULAR CABLE. THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED THROUGH THIS ANALYSIS. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE RECORDS REVEALED THAT THE HEARTMATE 3 SYSTEM CONTROLLER WAS MANUFACTURED IN ACCORDANCE WITH MANUFACTURING AND QA SPECIFICATIONS. THE MODULAR CABLE WAS SHIPPED TO THE CUSTOMER. HEARTMATE 3 INSTRUCTIONS FOR USE SECTION 7, ENTITLED ¿ALARMS AND TROUBLESHOOTING¿, AND HEARTMATE 3 PATIENT HANDBOOK SECTION 5, ENTITLED ¿ALARMS AND TROUBLESHOOTING¿ COVERS ALL ALARMS (VISUAL AND AUDIBLE), INCLUDING THE DRIVELINE POWER FAULT AND BACKUP BATTERY FAULT ALARM CONDITIONS, AND THE ACTIONS TO TAKE IF THE ALARM CANNOT BE RESOLVED. THE SUBSECTION ENTITLED ¿WHAT NOT TO DO: DRIVELINE AND CABLES¿ DESCRIBES CHECKING THE DRIVELINE CABLE FOR TWISTING, KINKING, OR BENDING WHICH COULD CAUSE DAMAGE TO THE WIRES INSIDE, EVEN IF EXTERNAL DAMAGE IS NOT VISIBLE. DAMAGE TO THE DRIVELINE OR CABLES COULD CAUSE THE LEFT VENTRICULAR ASSIST DEVICE TO STOP. HEARTMATE 3 INSTRUCTIONS FOR USE SECTION 8, ENTITLED ¿EQUIPMENT STORAGE AND CARE¿, AND HEARTMATE 3 PATIENT HANDBOOK SECTION 6, ENTITLED ¿CARING FOR EQUIPMENT¿, EXPLAIN HOW TO PROPERLY CARE FOR THE DRIVELINE CABLE, AND THE SYSTEM CONTROLLER. HEARTMATE 3 PATIENT HANDBOOK SECTION 10, ENTITLED ¿SAFETY CHECKLISTS¿, AND HEARTMATE 3 INSTRUCTIONS FOR USE SECTION F, ENTITLED ¿SAFETY CHECKLIST¿, PROVIDE CHECKLISTS TO ASSIST THE PATIENT IN PERFORMING ROUTINE MAINTENANCE OF HEARTMATE 3 LVAD, INCLUDING INSPECTING THE SYSTEM CONTROLLER, THE SYSTEM CONTROLLER POWER CABLES, AND THE DRIVELINE FOR DAMAGE. THE HEARTMATE 3 PATIENT HANDBOOK CAUTIONS THE USER TO CALL THEIR HOSPITAL CONTACTS IF THEY THINK, FOR ANY REASON, ANY PORTION OF THEIR EQUIPMENT IS NOT FUNCTIONING AS USUAL, IS BROKEN, OR THEY ARE UNCOMFORTABLE WITH THE OPERATION OF THE EQUIPMENT. NO FURTHER INFORMATION WAS PROVIDED. THE MANUFACTURER IS CLOSING THE FILE ON THIS EVENT.

Description of Event or Problem · 0

RELATED MANUFACTURER REPORT NUMBER: 2916596-2023-05924. IT WAS REPORTED THAT THE PATIENT HAD A DRIVELINE POWER FAULT. THE MODULAR CABLE WAS ALSO TWISTED. THEIR SYSTEM CONTROLLER AND MODULAR CABLE WERE EXCHANGED. THE ALARMS DID NOT RETURN FOLLOWING THE EQUIPMENT EXCHANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
302003 HEARTMATE 3 VAD MODULAR CABLE Ventricular (assist) bypass DSQ THORATEC CORPORATION 106525US 8140086 00813024013181

Patients

Seq Age Sex Outcome Treatment
1 61 YR Male