FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 17608859 · Received August 24, 2023

Report

Report Number
3002808148-2023-08863
Event Type
Malfunction
Date Received
August 24, 2023
Report Date
October 12, 2023
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170363498
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION WAS CONFIRMED. BASED ON THE RESULTS OF THE INVESTIGATION, THE POWER WAS TURNED OFF AFTER START-UP DUE TO FAULTY PRINTED CIRCUIT (G1) BOARD. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 6 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION AND THE CUSTOMER'S ALLEGATION WAS CONFIRMED. THE EVALUATION FOUND THE POWER TURNS OFF DURING USE. FURTHERMORE, THE FOLLOWING ADDITIONAL ISSUES WERE IDENTIFIED DURING INSPECTION: THE SCREEN FRAME IS CRACKED, AND THE POWER SWITCH BRACKET IS RATTLING, THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, THE HIGH DEFINITION LCD MONITOR SELF-REBOOTS AND DISPLAYS NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308188 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET SHIRAKAWA OLYMPUS CO., LTD. OEV262H 04953170363498

Patients

Seq Age Sex Outcome Treatment
1 Unknown