ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2010-01323
- Event Type
- Injury
- Date Received
- July 15, 2010
- Date of Event
- May 1, 2010
- Report Date
- June 16, 2010
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): EVAL RESULTS: GRAFT OCCLUSION. SEVERE VESSEL TORTUOSITY AND SEVERE VESSEL CALCIFICATION.
A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 2 MONTHS AGO. ANEURYSM MORPHOLOGY WAS NOT REPORTED. VESSELS HAD SEVERE TORTUOSITY AND SEVERE CALCIFICATION. THE EXTERNAL ILIAC ARTERY WAS 9 MM IN DIAMETER. IT WAS REPORTED THAT AN ANEURX CONTRALATERAL LIMB AND A TALENT BIFURCATED STENT GRAFT (MFR REPORT # 2953200-2010-01322) WERE IMPLANTED WITH THE LIMBS UNCROSSED. APPROX 1 WEEK POST-IMPLANTATION, THE PT PRESENTED WITH LEG PAIN, AND THE CT DEMONSTRATED THAT THE IPSILATERAL LIMB OF THE BIFURCATED STENT GRAFT AND THE CONTRALATERAL ANEURX LIMB WERE BOTH KINKED AND OCCLUDED. THE KINK AND OCCLUSION WERE WHERE THE TWO STENT GRAFTS JOINED, AND MAY HAVE BEEN CAUSED BY THE SEVERE VESSEL TORTUOSITY. THE PHYSICIAN ELECTED TO PERFORM A FEMORAL-FEMORAL BYPASS, WHICH SUCCESSFULLY RESOLVED THE OCCLUSION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00441112 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Required Intervention |