FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 1760732 · Received July 15, 2010

Report

Report Number
2953200-2010-01322
Event Type
Injury
Date Received
July 15, 2010
Date of Event
May 1, 2010
Report Date
June 16, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS: GRAFT OCCLUSION. SEVERE VESSEL TORTUOSITY AND SEVERE VESSEL CALCIFICATION.

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 2 MONTHS AGO. ANEURYSM MORPHOLOGY WAS NOT REPORTED. VESSELS HAD SEVERE TORTUOSITY AND SEVERE CALCIFICATION. THE EXTERNAL ILIAC ARTERY WAS 9 MM IN DIAMETER. IT WAS REPORTED THAT AN ANEURX CONTRALATERAL LIMB (MFR REPORT # 2953200-2010-01323) AND A TALENT BIFURCATED STENT GRAFT WERE IMPLANTED WITH THE LIMBS UNCROSSED. APPROX 1 WEEK POST-IMPLANTATION, THE PT PRESENTED WITH LEG PAIN, AND THE CT DEMONSTRATED THAT THE IPSILATERAL LIMB OF THE BIFURCATED STENT GRAFT AND THE CONTRALATERAL ANEURX LIMB WERE BOTH KINKED AND OCCLUDED. THE KINK AND OCCLUSION WERE WHERE THE TWO STENT GRAFTS JOINED, AND MAY HAVE BEEN CAUSED BY THE SEVERE VESSEL TORTUOSITY. THE PHYSICIAN ELECTED TO PERFORM A FEMORAL-FEMORAL BYPASS, WHICH SUCCESSFULLY RESOLVED THE OCCLUSION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00475161

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention