DISTRACTION
Report
- Report Number
- 9610905-2023-00042
- Event Type
- Injury
- Date Received
- August 23, 2023
- Date of Event
- July 26, 2023
- Report Date
- May 15, 2024
- Manufacturer
- KLS MARTIN SE & CO. KG
- Product Code
- MQN
- UDI-DI
- 00888118113405
- PMA / PMN Number
- K010139
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
AN INVESTIGATION WAS PERFORMED ON THE BASIS OF COMPLAINT STATISTICS AS NO DEVICE WAS RETURNED FOR EVALUATION. THE COMPLAINT PERCENTAGE WAS CALCULATED, AND IT IS DETERMINED THAT THE COMPLAINT PERCENTAGE FALLS WITHIN THE DESIGN RISK LIMITS ADHERED TO AT KLS SE. DURING THE INVESTIGATION THE PRODUCT LOT NUMBER IDENTIFIED WAS REVIEWED IN THE DEVICE HISTORY RECORDS. THE DHR REVIEW SHOWED NO DISCREPANCIES OR ANOMALIES. THE FAILURE ROOT CAUSE CANNOT BE DETERMINED DUE TO NO DEVICE BEING RETURNED. IF FURTHER INFORMATION IS OBTAINED THAT MIGHT ADD VALUE TO THE CONTENTS OF THE INVESTIGATION REPORT, AN ADDITIONAL FOLLOW-UP REPORT WILL BE SUBMITTED. CORRECTIONS: D3 ESTABLISHMENT NAME - KLS MARTIN SE & CO. KG. G1 CONTACT OFFICE ESTABLISHMENT NAME - KLS-MARTIN L.P. G1 CONTACT OFFICE MANUFACTURING SITE NAME AND ADDRESS - KLS MARTIN SE & CO. KG. KOLBINGER STRASSE 10. MUEHLHEIM/DONAU, DEU 78570, DEU.
IT WAS REPORTED THAT SCREWS SECURING A DISTRACTOR BECAME LOOSE FROM BONE. THE PHYSICIAN REMOVED THE SCREWS AND REPLACED THEM WITH THE SAME SIZE AND MODEL SCREWS WHILE LEAVING THE DISTRACTOR IN PLACE. THE SCREWS WERE NOTED TO BE INTACT AND UNDAMAGED UPON EXPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1524790 | DISTRACTION | DISTRACTOR SCREW | MQN | KLS MARTIN SE & CO. KG | 50-347-07-61 | 33507577 | 00888118113405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1 MO | Female | Other |