LIFEVEST WCD 4000 SYSTEM
MDR report key: 1760660
·
Received July 13, 2010
Report
- Report Number
- 3002158293-2010-00725
- Event Type
- Malfunction
- Date Received
- July 13, 2010
- Date of Event
- June 6, 2010
- Report Date
- July 13, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE ALARMS TO CONNECT THE ELECTRODE BELT WAS A BROKEN TRUNK CABLE CONNECTOR. THE ROOT CAUSE OF THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT HE WAS GETTING A CODE 204 ON HIS MONITOR SCREEN. THE PT'S ELECTRODE BELT WAS REPLACED.