FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 1760658
·
Received July 13, 2010
Report
- Report Number
- 3002158293-2010-00723
- Event Type
- Malfunction
- Date Received
- July 13, 2010
- Date of Event
- December 17, 2009
- Report Date
- July 12, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY FAULTS ON DOWNLOAD) WAS CONFIRMED. THE CAUSE OF THE BATTERY NOT FULLY CHARGING WAS A BROKEN WHITE WIRE ON THE BATTERY PCA BOARD WHERE THE WIRE ATTACHES TO THE BATTERY CELLS. THE ROOT CAUSE OF THE BROKEN WIRE HAS NOT BEEN DETERMINED BUT MAY HAVE BEEN CAUSED BY A SEVERE SHOCK FROM DROPPING THE PACK. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK .
Description of Event or Problem · 1
DURING A REVIEW OF A (B)(6) FEMALE PT'S DOWNLOAD ZOLL LIFECOR CUSTOMER SUPPORT IDENTIFIED SEVERAL "BATTERY CHARGER FAULT" FLAGS IN THE DATA. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |