FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 1760614 · Received July 12, 2010

Report

Report Number
3002158293-2010-00720
Event Type
Malfunction
Date Received
July 12, 2010
Date of Event
June 15, 2010
Report Date
July 12, 2010
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE ALARMS TO CONNECT THE ELECTRODE BELT WAS A BROKEN TRUNK CABLE CONNECTOR. THE ROOT CAUSE OF THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT HE IS RECEIVING ALARMS TO CONNECT THE ELECTRODE BELT. THE PT HAD ALSO REPORTED THAT THE CONNECTOR WAS DAMAGED. THE PT'S ELECTRODE BELT WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR