FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM (COILTRAC)

MDR report key: 1760597 · Received July 14, 2010

Report

Report Number
2953200-2010-01313
Event Type
Injury
Date Received
July 14, 2010
Date of Event
June 15, 2010
Report Date
June 15, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL, RESULTS: (ENDOLEAK). (LIMB PLACEMENT WAS SHORT IN THE ILIAC ARTERY). CONCLUSION: (LIMB PLACEMENT WAS SHORT IN THE ILIAC ARTERY).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX 21 MONTHS AGO. ANEURYSM AND VESSEL MORPHOLOGY AT THE TIME OF IMPLANT WERE NOT REPORTED. AT THE TIME OF THE INITIAL IMPLANT, THE CONTRALATERAL AND IPSILATERAL LIMBS WERE BOTH PLACED INTO THE COMMON ILIAC ARTERIES BY ONLY ABOUT 1.5 CM. IT WAS REPORTED THAT APPROXIMATELY 1 MONTH AGO, THE PT PRESENTED EMERGENTLY WITH ABDOMINAL AND BACK PAIN. AT THE TIME OF THE EVENT, THE ANEURYSM WAS 5.5-6 CM IN DIAMETER, AND THE AORTIC NECK AND VESSELS WERE UNREMARKABLE. A DISTAL TYPE I ENDOLEAK WAS NOTED ON THE LEFT/CONTRALATERAL SIDE. THE PHYSICIAN ELECTED TO INTERVENE BY IMPLANTING AN ANEURX EXTENSION GRAFT DISTALLY. ALSO, BECAUSE THE IPSILATERAL LIMB (MFR REPORT # 2953200-2010-01314) WAS SHORT IN THE ILIAC ARTERY AT THE TIME OF IMPLANT, THE PHYSICIAN ADDED AN ANEURX EXTENSION AS A PROPHYLACTIC MEASURE, EVEN THOUGH NO ENDOLEAK WAS NOTED. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM (COILTRAC) MIH MEDTRONIC CARDIOVASCULAR NA V00125691

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention