FDA Adverse Event Other Summary report: N

OLYMPUS URETERO-RENO VIDEOSCOPE

MDR report key: 1760587 · Received July 13, 2010

Report

Report Number
8010047-2010-00134
Event Type
Other
Date Received
July 13, 2010
Date of Event
June 18, 2010
Report Date
June 21, 2010
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
FGB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS FOR EVALUATION. THE EVALUATION FOUND THE BENDING COVER LEAKING DUE TO A TORN BENDING COVER AT THE DISTAL END. METAL BRAIDS IN THE BENDING SECTION WERE BUNCHED UP, BROKEN, AND SOME WERE OBSERVED TO BE PROTRUDING OUT FROM THE BENDING COVER. THE DISTAL END COVER WAS NOTED BURNT AND MELTED AT THE CHANNEL PIPE OPENING. THE DEVICE HAS BEEN FORWARDED TO THE ORIGINAL EQUIPMENT MANUFACTURER FOR FURTHER INVESTIGATION. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION. IF ADDITIONAL AND SIGNIFICANT INFORMATION BECOMES AVAILABLE A SUPPLEMENTAL REPORT WILL FOLLOW. (B)(4).

Description of Event or Problem · 1

DURING A THERAPEUTIC URETEROSCOPY PROCEDURE, THE PHYSICIAN REPORTEDLY ATTEMPTED TO PASS THE URETERO-RENO VIDEOSCOPE THROUGH AN 11 FRENCH INNER DIAMETER ACCESS SHEATH. AS THE USERS WERE ATTEMPTING TO MOVE THROUGH A TIGHT SPOT, THE BENDING SECTION REPORTEDLY BECAME STUCK AND WAS DAMAGED. THERE WAS NO ALLEGATION OF PARTS/DEBRIS FALLING IN THE PATIENT. THE USERS REPORTEDLY COULD NOT REMOVE THE SCOPE. THE SHEATH AND VIDEOSCOPE WERE REMOVED SIMULTANEOUSLY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING AN UNSPECIFIED FLEXIBLE ENDOSCOPE AND NO ACCESS SHEATH. THERE WAS NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPUS URETERO-RENO VIDEOSCOPE URETERO-RENO VIDEOSCOPE FGB OLYMPUS MEDICAL SYSTEMS CORPORATION URF-V NA

Patients

Seq Age Sex Outcome Treatment
1 (13FR OUTER DIAMETER)| MICROVASIVE ACCESS SHEATH| UNKNOWN MODEL, 11FR INNER DIAMETER| UNKNOWN MODEL, 11FR INNER DIAMETER| MICROVASIVE ACCESS SHEATH| (13FR OUTER DIAMETER)