FDA Adverse Event
Injury
Summary report: N
PUMP MMT-722NAP PRDGM INS V2.2 PL EN
MDR report key: 1760438
·
Received July 13, 2010
Report
- Report Number
- 3004209178-2010-81976
- Event Type
- Injury
- Date Received
- July 13, 2010
- Date of Event
- June 27, 2010
- Report Date
- June 27, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETES KETOACIDOSIS. THE BLOOD GLUCOSE READING WAS OVER 600MG/DL. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING, TIME, BASALS, DAILY TOTALS AND BOLUS HISTORY MATCHED. RAN A FIXED PRIME AND HIGH PRESSURE TESTS AND PASSED. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722NAP PRDGM INS V2.2 PL EN | INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR | Hospitalization |