BINAXNOW COVID-19 AG SELF TEST 2CT
Report
- Report Number
- 1221359-2023-01361
- Event Type
- Malfunction
- Date Received
- August 23, 2023
- Date of Event
- August 2, 2023
- Report Date
- September 15, 2023
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA210264
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
D4: UDI - (B)(4). TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 175523 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160/ LOT 175523, TEST BASE PART NUMBER 195-430H/ LOT 172269. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 177523 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE. H3 OTHER TEXT : SINGLE-USE: DEVICE DISCARDED.
THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. H3 OTHER TEXT : SINGLE-USE: DEVICE DISCARDED.
THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH A BINAXNOW COVID-19 ANTIGEN SELF-TEST ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING WAS NOT PERFORMED. CONSUMER PERFORMED AN ANTIGEN SELF-TEST WITH AN INBIOS TEST ON THE SAME DAY AND GENERATED A NEGATIVE RESULT. THE CONSUMER CONFIRMED THERE WAS NO HARM DUE TO THE TEST RESULT. ADDITIONALLY, THE CONSUMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THEIR TREATMENT.
THE CONSUMER REPORTED A FALSE POSITIVE RESULT WITH A BINAXNOW COVID-19 ANTIGEN SELF-TEST ON (B)(6) 2023 ON A NASAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING WAS NOT PERFORMED. CONSUMER PERFORMED AN ANTIGEN SELF-TEST WITH AN INBIOS TEST ON THE SAME DAY AND GENERATED A NEGATIVE RESULT. THE CONSUMER CONFIRMED THERE WAS NO HARM DUE TO THE TEST RESULT. ADDITIONALLY, THE CONSUMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THEIR TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1525007 | BINAXNOW COVID-19 AG SELF TEST 2CT | CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM. | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 175523 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Female |