FDA Adverse Event
Malfunction
Summary report: N
MONOPOLAR CURVED SCISSORS INSTRUMENT
MDR report key: 1760253
·
Received July 16, 2010
Report
- Report Number
- 2955842-2010-00321
- Event Type
- Malfunction
- Date Received
- July 16, 2010
- Date of Event
- March 23, 2010
- Report Date
- May 10, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. PER THE INFORMATION PROVIDED BY THE HOSPITAL, THE BROKEN PIECE WAS SUCCESSFULLY RETRIEVED FROM THE PATIENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE TIP OF THE MONOPOLAR CURVED SHEARS (MCS) INSTRUMENT BROKE OFF AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS RETRIEVED AND THE PLANNED PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-09 | M10100107 178 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYS, MCS TIP COVER ACCESSORY & AN ES |