FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CXE VOLUMETRICINFUSION PUMP

MDR report key: 1760244 · Received July 16, 2010

Report

Report Number
6000001-2010-01726
Event Type
Malfunction
Date Received
July 16, 2010
Date of Event
July 9, 2010
Report Date
July 9, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION MDQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE GREY COLORED HYDROCANISTER LID OF THE UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER CRACKED. NO INJURY WAS REPORTED.

Description of Event or Problem · 1

THE SERVICE TECHNICIAN FOUND A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 810:11 CAUSED BY AIL (AIR IN LINE) OUT OF CALIBRATION AND HAD TO BE RECALIBRATED. THIS EVENT OCCURRED DURING SERVICE. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 6.13.90, CATEGORIZED AS REMEDIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CXE VOLUMETRICINFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1