FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 1760234 · Received July 16, 2010

Report

Report Number
2955842-2010-00324
Event Type
Malfunction
Date Received
July 16, 2010
Date of Event
May 5, 2010
Report Date
May 15, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Removal / Correction Number
2955842-051613-005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED FOR RETROSPECTIVE ACTIVITY PERFORMED RELATING TO FIELD ACTION NUMBER 2955842-051613-005 TO INVESTIGATE MICRO-CRACKS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT; HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESES MICRO-CRACKS IS CONFINED TO 2 CM OF THE DISTAL END OF THE INSTRUMENT SHAFT. THIS INSTRUMENT WAS INSPECTED AS PART OF THIS RETROSPECTIVE ACTIVITY AND FOUND TO CONTAIN CRACKS ON THE INSTRUMENT MAIN TUBE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, WHILE COAGULATING TISSUE, THE GRIP ON THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT BROKE AND A PIECE FROM THE INSTRUMENT FELL INTO THE PATIENT. THE BROKEN PIECE WAS NOT RETRIEVED FROM THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A REPLACEMENT INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-09 M10100112 731

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYS., MCS TIP COVER ACCESSORY & AN ESU