MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2010-00324
- Event Type
- Malfunction
- Date Received
- July 16, 2010
- Date of Event
- May 5, 2010
- Report Date
- May 15, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Removal / Correction Number
- 2955842-051613-005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION, THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED.
THIS MDR IS BEING SUBMITTED FOR RETROSPECTIVE ACTIVITY PERFORMED RELATING TO FIELD ACTION NUMBER 2955842-051613-005 TO INVESTIGATE MICRO-CRACKS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT; HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESES MICRO-CRACKS IS CONFINED TO 2 CM OF THE DISTAL END OF THE INSTRUMENT SHAFT. THIS INSTRUMENT WAS INSPECTED AS PART OF THIS RETROSPECTIVE ACTIVITY AND FOUND TO CONTAIN CRACKS ON THE INSTRUMENT MAIN TUBE.
IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, WHILE COAGULATING TISSUE, THE GRIP ON THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT BROKE AND A PIECE FROM THE INSTRUMENT FELL INTO THE PATIENT. THE BROKEN PIECE WAS NOT RETRIEVED FROM THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH A REPLACEMENT INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-09 | M10100112 731 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI S SYS., MCS TIP COVER ACCESSORY & AN ESU |