PATELLA CALIPER
Report
- Report Number
- 2249697-2010-00823
- Event Type
- Malfunction
- Date Received
- July 12, 2010
- Date of Event
- June 16, 2010
- Report Date
- June 16, 2010
- Manufacturer
- STRYKER ORTHOPAEDICS MAHWAH
- Product Code
- LXH
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. THIS IS THE SAME PATIENT/EVENT AS: MFR # 2249697-2010-00818, MFR # 2249697-2010-00819, MFR # 2249697-2010-00820, MFR # 2249697-2010-00821, MFR # 2249697-2010-00822, MFR # 2249697-2010-00824, MFR # 2249697-2010-00825, MFR # 2249697-2010-00826.
IT WAS REPORTED THAT "ALL THESE ITEMS, WITH THE GREEN HANDLE, THE GREEN HANDLE IS DISINTEGRATING. (ALMOST LIKE THEY ARE MELTING, RUB HAND ON IT THE GREEN COMES OFF IN YOUR HAND.) REWASHED AND RESTERILIZED AND IT IS STILL HAPPENING. NOT SURE IF IT IS A PROCESSING ISSUE. AUTOCLAVE GOES TO 270 DEGREES FOR 50 MINUTES AND THEY USE A STERIS CLEANING AGENT. SEVERAL DIFFERENT TRAYS, ALL IN ONE LOAD, WASHED AND PROCESSED AT THE SAME TIME."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PATELLA CALIPER | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS MAHWAH | NA | ABZK25 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Other |