FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1760094 · Received July 9, 2010

Report

Report Number
2531779-2010-00597
Event Type
Malfunction
Date Received
July 9, 2010
Date of Event
May 4, 2010
Report Date
June 14, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED SEVERAL LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE.

Description of Event or Problem · 1

EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1