FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 17600647
·
Received August 23, 2023
Report
- Report Number
- 9610877-2023-57755
- Event Type
- Malfunction
- Date Received
- August 23, 2023
- Date of Event
- August 8, 2023
- Report Date
- August 23, 2023
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EOB
- UDI-DI
- 04961333070589
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FNL-10RP3 IS AVAILABLE IN THE USA WITH A 510K NUMBER K951196. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OCULAR FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OCULAR. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE(FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1405385 | PENTAX | FIBER NASO PHARYNGO LARYNGOSCOPE | EOB | HOYA CORPORATION PENTAX TOKYO OFFICE | FNL-10RBS | 04961333070589 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |