FDA Adverse Event Malfunction Summary report: N

COLLEAGUE TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1760012 · Received July 16, 2010

Report

Report Number
6000001-2010-01700
Event Type
Malfunction
Date Received
July 16, 2010
Date of Event
May 13, 2010
Report Date
June 2, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Removal / Correction Number
6000001-3/15/05-007-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE FOR THIS ISSUE IS CURRENTLY BEING INVESTIGATED THROUGH CAPA (CORRECTIVE AND PREVENTIVE ACTION) INVESTIGATION (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND EVALUATED BY BAXTER. THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE IS THE PHM (PUMPHEAD MODULE). THE PHM WAS REPLACED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER (B)(4) TO REPORT A COLLEAGUE INFUSION PUMP THAT HAD A DAMAGED BATTERY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. DURING BAXTER'S REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT THERE WAS A BATTERY DEPLETED SET ALARM THAT HAD OCCURRED DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN FOR THIS DEVICE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH AN 810:04 FAILURE CODE. THE EVENT WAS REPORTED TO HAVE OCCURRED DURING PROGRAMMING/SET-UP IN THE GENERAL PATIENT WARD. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY, ADVERSE EVENT OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.03.00 WHICH IS CATEGORIZED AS UNREMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE.